Iris Diagnostics: IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.
- Recall number
- Z-2124-2012
- Recalling firm
- Iris Diagnostics
- Product
- IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.
- Status
- Terminated
- Initiated
- 2012-07-09
- Posted
- 2012-08-01
- Terminated
- 2014-01-27
- Reason
- The firm recalled after it received complaints of IRISpec CA/CB two part Urine Chemistry Control, 101131-12, where the nitrite results read negative for the part B control, which should be positive.
- Root cause
- Nonconforming Material/Component
- Action
- The firm initiated the recall on 07/09/12 by sending out a Medical Device Recall notification to customers via e-mail. The recall notification identified the affected product, reason for recall, and instructions on the recall. Customers were to confirm how many sets of unopened or partially opened IRISpec CA/CB that was in their inventory, dispose of the affected product, and provide written confirmation of the product's disposition. The letter also stated that Iris Diagnostics would replace any unused inventory that was returned at no charge to the customer. Contact information for the recall was also provided. The communications also included a fax response sheet for return/replacement.
- Quantity
- 862
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Iris Diagnostics part number 800-0074, lot number 131-12.
- 510(k) numbers
- ["K945913"]
- PMA numbers
- not on file
- Product code
- JJW
- Firm FEI
- 2023446
- Firm city
- Chatsworth
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28