Skip to the content

FDA Medical Device Recalls (openFDA)

Iris Diagnostics: IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.

Recall number
Z-2124-2012
Recalling firm
Iris Diagnostics
Product
IRISpec" CA/CB two-part urine chemistry control twin set, Iris Diagnostics part number 800-0074. Intended Use: IRISpec CA/CB is an assayed QC material for monitoring of the urine chemistry analytes and devices as listed on the package insert. For in vitro diagnostic (IVD) use only.
Status
Terminated
Initiated
2012-07-09
Posted
2012-08-01
Terminated
2014-01-27
Reason
The firm recalled after it received complaints of IRISpec CA/CB two part Urine Chemistry Control, 101131-12, where the nitrite results read negative for the part B control, which should be positive.
Root cause
Nonconforming Material/Component
Action
The firm initiated the recall on 07/09/12 by sending out a Medical Device Recall notification to customers via e-mail. The recall notification identified the affected product, reason for recall, and instructions on the recall. Customers were to confirm how many sets of unopened or partially opened IRISpec CA/CB that was in their inventory, dispose of the affected product, and provide written confirmation of the product's disposition. The letter also stated that Iris Diagnostics would replace any unused inventory that was returned at no charge to the customer. Contact information for the recall was also provided. The communications also included a fax response sheet for return/replacement.
Quantity
862
Distribution
Nationwide Distribution
Lot, serial or model codes
Iris Diagnostics part number 800-0074, lot number 131-12.
510(k) numbers
["K945913"]
PMA numbers
not on file
Product code
JJW
Firm FEI
2023446
Firm city
Chatsworth
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register