Skip to the content

FDA Medical Device Recalls (openFDA)

Davol, Inc., Sub. C. R. Bard, Inc.: BardVentralex" Hernia Patch Medium Circle with Strap Diameter 6.4cm /2.5",\ Catalog #0010302, lot number HUTL1283. The BardVentralex" Hernia Patch is a self-expanding, non-absorbable, sterile prosthesis. The device classification name is "Mesh, Surgical, Polymeric". It is intended for use in all forms of hernia repair requiring reinforcement with a non-absorbable support material

Recall number
Z-2124-2011
Recalling firm
Davol, Inc., Sub. C. R. Bard, Inc.
Product
BardVentralex" Hernia Patch Medium Circle with Strap Diameter 6.4cm /2.5",\ Catalog #0010302, lot number HUTL1283. The BardVentralex" Hernia Patch is a self-expanding, non-absorbable, sterile prosthesis. The device classification name is "Mesh, Surgical, Polymeric". It is intended for use in all forms of hernia repair requiring reinforcement with a non-absorbable support material
Status
Terminated
Initiated
2011-04-11
Posted
2011-05-02
Terminated
2011-10-27
Reason
Mislabeled: Bard Medium Ventralex Hernia Patch 6.4cm /2.5" packaged in a pre-printed carton identified for a Bard Ventralex Small Circle 4.3cm /1.7"
Root cause
Labeling mix-ups
Action
Davol sent an Urgent: Medical Device Recall letter on April 11, 2011, via FedeX Priority Overnight to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and remove and return the recalled lot. Distributors were requested to notify their accounts. For questions contact the Davol Customer Service Department at 1-800-556-6275 or C. R. Bard's Medical Services & Support Department at 1-800-562-0027.
Quantity
146 units
Distribution
Nationwide Distribution - USA including AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, MA, MD, MI, MO, MT, NC, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, VA, WA, and WV.
Lot, serial or model codes
Lot number HUTL1283
510(k) numbers
["K021736"]
PMA numbers
not on file
Product code
FTL
Firm FEI
1213643
Firm city
Warwick
Firm state
RI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register