FDA Medical Device Recalls (openFDA)
Orthofix, Inc: Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a)D 3.2MM X 10MM (002-D-32010) b) D 3.2MM X 12MM (002-D-32012) c) D 3.2MM X 14MM (002-D-32014) d) D 3.2MM X 16MM (002-D-32016) e) D 3.2MM X 18MM (002-D-32018) f) D 3.2MM X 20MM (002-D-32020) g) D 3.2MM X 22MM (002-D-32022) h) D 3.2MM X 24MM (002-D-32024) i) D 3.2MM X 26MM (002-D-32026) j) D 3.2MM X 28MM (002-D-32028) k) D 3.2MM X 30MM (002-D-32030) l) D 3.2MM X 32MM (002-D-32032) m) D 3.2MM X 34MM (002-D-32034) n) D 3.2MM X 36MM (002-D-32036) o) D 3.2MM X 38MM (002-D-32038) p) D 3.2MM X 40MM (002-D-32040) q) D 3.2MM X 42MM (002-D-32042) r) D 3.2MM X 44MM (002-D-32044) s) D 3.2MM X 46MM (002-D-32046) t) D 3.2MM X 48MM (002-D-32048) u) D 3.2MM X 50MM (002-D-32050)
- Recall number
- Z-2123-2018
- Recalling firm
- Orthofix, Inc
- Product
- Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a)D 3.2MM X 10MM (002-D-32010) b) D 3.2MM X 12MM (002-D-32012) c) D 3.2MM X 14MM (002-D-32014) d) D 3.2MM X 16MM (002-D-32016) e) D 3.2MM X 18MM (002-D-32018) f) D 3.2MM X 20MM (002-D-32020) g) D 3.2MM X 22MM (002-D-32022) h) D 3.2MM X 24MM (002-D-32024) i) D 3.2MM X 26MM (002-D-32026) j) D 3.2MM X 28MM (002-D-32028) k) D 3.2MM X 30MM (002-D-32030) l) D 3.2MM X 32MM (002-D-32032) m) D 3.2MM X 34MM (002-D-32034) n) D 3.2MM X 36MM (002-D-32036) o) D 3.2MM X 38MM (002-D-32038) p) D 3.2MM X 40MM (002-D-32040) q) D 3.2MM X 42MM (002-D-32042) r) D 3.2MM X 44MM (002-D-32044) s) D 3.2MM X 46MM (002-D-32046) t) D 3.2MM X 48MM (002-D-32048) u) D 3.2MM X 50MM (002-D-32050)
- Status
- Terminated
- Initiated
- 2018-04-06
- Posted
- 2018-06-07
- Terminated
- 2021-06-30
- Reason
- Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Root cause
- Device Design
- Action
- On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
- Quantity
- 1206
- Distribution
- U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
- Lot, serial or model codes
- (UDI)/Lot: a) (18056099645959) B1082246; b) (18056099645966) B1075836; c) (18056099645973) B1084806; d) (18056099645980) B1089892; e) (18056099645997) B1075845; f) (18056099646000) B1084797; g) (18056099646017) B1084792; h) (18056099646024) B1084799; i) (18056099646031) B1083858; j) (18056099646048) B1083859; k) (18056099646055) B1089891, B1195735; l) (18056099646062) B1195736, B1089882; m)(18056099646079) B1204992, B1083850; n) (18056099646086) B1083848, B1195699; o) (18056099646093) B1195732, B1089896; p) (18056099646109) B1204991, B1083642; q) (18056099646116) B1195739, B1083851; r) (18056099646123) B1204988, B1083637; s) (18056099646130) B1195733, B1089898; t) (18056099646147) B1195744, B1083838; u) (18056099646154) B1195730, B1071599, B1083853
- 510(k) numbers
- ["K140876"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 2183449
- Firm city
- Lewisville
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28