Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Intended to be used to acquire, store, print, transfer, and archive clinical information.

Recall number
Z-2123-2017
Recalling firm
Merge Healthcare, Inc.
Product
Merge Cardio software. The firm name on the label is Merge Healthcare, Hartland, WI. Intended to be used to acquire, store, print, transfer, and archive clinical information.
Status
Terminated
Initiated
2016-04-13
Posted
2017-05-22
Terminated
2017-07-17
Reason
Fetal patient report was automatically pulling prior measurement data for a prior fetus since the fetal study is based on the mother's MRN, resulting in the wrong fetal measurements getting referenced.
Root cause
Software design
Action
The recalling firm issued a "Medical Device Correction" letter dated 4/13/2016 via e-mail. The letter identified the affected product, issue involved, potential harm, a work around, as well as the actions to be performed by both the customer and firm. Merge Technical Support will contact customers to discuss the fix. Customers are to ensure that all users are provided the notice. If the affected product was further distributed, those customers should be identified and notified of the issue. A copy of the notice may be provided to those customers. If there are additional questions, customers can e-mail recall@merge.com.
Quantity
30 sites have the potentially affected software versions
Distribution
Distributed to the states of AZ, CT, FL, GA, IL, IN, LA, MI, NY, NC, OH, OK, TX, and VT.
Lot, serial or model codes
Versions 10.0, 10.0.1, 10.1, 10.1.1,10.1.1.1, and 10.1.2.
510(k) numbers
["K051649"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register