GE Healthcare: GE Healthcare, Precision MPi fluoroscopic imaging systems.

Recall number
Z-2123-2015
Recalling firm
GE Healthcare
Product
GE Healthcare, Precision MPi fluoroscopic imaging systems.
Status
Terminated
Initiated
2015-06-12
Posted
2015-07-20
Terminated
2017-08-03
Reason
A reported incident of a monitor that fell from the suspension. A fall of a monitor could result in an injury to a patient or operator. Recall affects the single and dual LCD monitor suspensions used on Precision 500D Classical R & F system, Precision RXi 23A/32A Analog System, Precision RXi Digital system and Precision MPi fluoroscopic imaging systems.
Root cause
Nonconforming Material/Component
Action
GE Healthcare sent an "Urgent Medical Device Correction" letter dated June 12, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Perform a visual check before and after each use for the presence of the four (4) mounting screws as shown in the pictures below. Check the secureness of these screws by hand. If these screws are present and secure, you may continue use of the system. If any of these screws are missing or can easily be turned by hand, follow the precautions below before you continue use with the monitor(s). 1. Position the monitor suspension in the most frequently used position and limit further movement as much as possible. 2. Clearly inform (e.g. through signage and verbal instructions) the users and other auxiliary personnel who may come in contact with the system to not move the monitor suspension once in position. 3. Contact your GE Healthcare representative to inform them of any loose or missing screws. For questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
788 (701 US; 87 OUS)
Distribution
Worldwide Distribution - US Nationwide Distribution including GU and DC., and to the countries of : Australia, Bulgaria, Canada, Chile, Dominican Republic, Germany, Ireland, Italy, Kuwait, Lebanon, Portugal, Romania, Russia, Saudi Arabia, Slovakia, Spain, Taiwan,Turkey, United Kingdom, Venezuela and Yemen.
Lot, serial or model codes
Mfg Lot or Serial # 00000000687MC4 00000001286MC4 00000004134MC3 00000003843MC0 00000003890MC1 00000004159MC0 00000004689MC6 00000001505MC7 00000003889MC3 00000003937MC0 00000002207MC9 00000003931MC3 00000001452MC2 00000003847MC1 00000004427MC1 00000004649MC0 00000003642MC6 00000001343MC3 00000004650MC8 00000004161MC6 00000001459MC7 00000001658MC4 00000001333MC4 00000003511MC3 00000001430MC8 00000001057MC9 00000001055MC3 00000001275MC7 00000001420MC9 00000001406MC8 00000001889MC5 00000001408MC4 00000003936MC2 00000001101MC5 00000001415MC9 00000001746MC7 00000001419MC1 00000004701MC9 0000000470FMC9 00000004700MC1 00000000952MC2 00000001302MC9 00000000921MC7
510(k) numbers
["K033486"]
PMA numbers
not on file
Product code
JAA
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register