Orthofix, Inc: Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 2.7MM X 8MM (002-E-27008), b) D 2.7MM X 10MM (002-E-27010); c) D 2.7MM X 12MM (002-E-27012); d) D 2.7MM X 14MM (002-E-27014); e) D 2.7MM X 16MM (002-E-27016); f) D 2.7MM X 18MM (002-E-27018); g) D 2.7MM X 20MM (002-E-27020); h) D 2.7MM X 22MM (002-E-27022); i) D 2.7MM X 24MM (002-E-27024); j) D 2.7MM X 26MM (002-E-27026); k) D 2.7MM X 28MM (002-E-27028); l) D 2.7MM X 30MM (002-E-27030); m) D 2.7MM X 32MM (002-E-27032); n) D 2.7MM X 34MM (002-E-27034); o) D 2.7MM X 36MM (002-E-27036); p) D 2.7MM X 38MM (002-E-27038); q) D 2.7MM X 40MM (002-E-27040)

Recall number
Z-2122-2018
Recalling firm
Orthofix, Inc
Product
Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) D 2.7MM X 8MM (002-E-27008), b) D 2.7MM X 10MM (002-E-27010); c) D 2.7MM X 12MM (002-E-27012); d) D 2.7MM X 14MM (002-E-27014); e) D 2.7MM X 16MM (002-E-27016); f) D 2.7MM X 18MM (002-E-27018); g) D 2.7MM X 20MM (002-E-27020); h) D 2.7MM X 22MM (002-E-27022); i) D 2.7MM X 24MM (002-E-27024); j) D 2.7MM X 26MM (002-E-27026); k) D 2.7MM X 28MM (002-E-27028); l) D 2.7MM X 30MM (002-E-27030); m) D 2.7MM X 32MM (002-E-27032); n) D 2.7MM X 34MM (002-E-27034); o) D 2.7MM X 36MM (002-E-27036); p) D 2.7MM X 38MM (002-E-27038); q) D 2.7MM X 40MM (002-E-27040)
Status
Terminated
Initiated
2018-04-06
Posted
2018-06-07
Terminated
2021-06-30
Reason
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Root cause
Device Design
Action
On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
Quantity
1302
Distribution
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Lot, serial or model codes
(UDI)/Lot: a) (18056099646338) B1075104; b) (18056099646345) B1071537, B1083847; c) (18056099646352) B1068544, B1075103; d) (18056099646369) B1075176; e) (18056099646376) B1075838; f) (18056099646383) B1069676, B1082189; g) (18056099646390) B1071602, B1082231; h) (18056099646406) B1069678, B1084787; i) (18056099646413) B1071672, B1082227, B1071672, B1075052; j) (18056099646420) B1082238; k) (18056099646437) B1084801; l) (18056099646444) B1082214, B1204276; m) (18056099646451) B1215467, B1089836; n) (18056099646468) B1215470, B1089877; o) (18056099646475) B1215466, B1089894; p) (18056099646482) B1195749, B1082234; q) (18056099646499) B1204583, B1089842
510(k) numbers
["K140876"]
PMA numbers
not on file
Product code
HRS
Firm FEI
2183449
Firm city
Lewisville
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register