FDA Medical Device Recalls (openFDA)
Baxter Healthcare Corp.: Baxter Flo-Gard 6201, Single Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8063 The Flo-Gard Volumetric Infusion Pumps are an electromechanical pump used for intravenous infusion of fluids at user-selected rates.
- Recall number
- Z-2122-2011
- Recalling firm
- Baxter Healthcare Corp.
- Product
- Baxter Flo-Gard 6201, Single Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8063 The Flo-Gard Volumetric Infusion Pumps are an electromechanical pump used for intravenous infusion of fluids at user-selected rates.
- Status
- Terminated
- Initiated
- 2011-03-24
- Posted
- 2011-04-29
- Terminated
- 2011-04-29
- Reason
- Pumps submitted for service may not have been properly serviced before being returned to the customer.
- Root cause
- Employee error
- Action
- Baxter visited the direct accounts starting on 3/24/11 informing them that the affected Flo-Gard pumps may have been released from the preventative maintenance with incorrect accuracy testing. The affected pumps were located, taken out of service and placed in quarantine on the same date. A technician not involved with the previous testing visited both accounts on 3/25/11 and performed preventative maintenance service of the three affected devices, and were found to be within specification.
- Quantity
- 3 units
- Distribution
- Puerto Rico
- Lot, serial or model codes
- Serial numbers 6100410FA, 8110687FA and 11021746FA
- 510(k) numbers
- ["K915522"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 1416980
- Firm city
- Round Lake
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28