Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Sofamor Danek USA Inc: Medtronic Catalyft LS Expandable Interbody System, Anterior Standalone Cage, REF 981027722, spinal implant

Recall number
Z-2121-2023
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
Medtronic Catalyft LS Expandable Interbody System, Anterior Standalone Cage, REF 981027722, spinal implant
Status
Open, Classified
Initiated
2023-06-06
Posted
2023-07-10
Terminated
not on file
Reason
There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
Root cause
Under Investigation by firm
Action
Medtronic issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 06/06/2023 by UPS 2-day delivery. The notice explained the problem , risk, and requested the following actions be taken: "" For purchased inventory not consumed during a procedure: o Identify and quarantine any unused impacted product(s). Refer to the table above for impacted products. o Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form. " For Medtronic owned product at your facility: o Your Medtronic Sales Representative will locate and return this product. " Complete the Customer Confirmation Form enclosed with this letter (even if you have no product to return), acknowledging that you have received this information. " This notice should be distributed to all others in your organization who should be aware, or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records."
Quantity
10 units
Distribution
US Nationwide distribution in the states of IN, WI, GA, OR, AR, OH, TN, TX, NC, OR, NE, MN, FL.
Lot, serial or model codes
UDI/DI 00763000465902, Lot Numbers: NM21K049
510(k) numbers
["K212653"]
PMA numbers
not on file
Product code
OVD
Firm FEI
1000477302
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register