FDA Medical Device Recalls (openFDA)
Orthofix, Inc: Non-Locking Screw, D 2.7MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Numbers): a) D 2.7MM X 8MM (002-D-27008); b) D 2.7MM X 10MM (002-D-27010); c) D 2.7MM X 12MM (002-D-27012); d) D 2.7MM X 14MM (002-D-27014); e) D 2.7MM X 16MM (002-D-27016); f) D 2.7MM X 18MM (002-D-27018); g) D 2.7MM X 20MM (002-D-27020); h) D 2.7MM X 22MM (002-D-27022); i) D 2.7MM X 24MM (002-D-27024); j) D 2.7MM X 26MM (002-D-27026); k) D 2.7MM X 28MM (002-D-27028); l) D 2.7MM X 30MM (002-D-27030); m) D 2.7MM X 32MM (002-D-27032); n) D 2.7MM X 34MM (002-D-27034); o) D 2.7MM X 36MM (002-D-27036); p) D 2.7MM X 38MM (002-D-27038); q) D 2.7MM X 40MM (002-D-27040)
- Recall number
- Z-2121-2018
- Recalling firm
- Orthofix, Inc
- Product
- Non-Locking Screw, D 2.7MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Numbers): a) D 2.7MM X 8MM (002-D-27008); b) D 2.7MM X 10MM (002-D-27010); c) D 2.7MM X 12MM (002-D-27012); d) D 2.7MM X 14MM (002-D-27014); e) D 2.7MM X 16MM (002-D-27016); f) D 2.7MM X 18MM (002-D-27018); g) D 2.7MM X 20MM (002-D-27020); h) D 2.7MM X 22MM (002-D-27022); i) D 2.7MM X 24MM (002-D-27024); j) D 2.7MM X 26MM (002-D-27026); k) D 2.7MM X 28MM (002-D-27028); l) D 2.7MM X 30MM (002-D-27030); m) D 2.7MM X 32MM (002-D-27032); n) D 2.7MM X 34MM (002-D-27034); o) D 2.7MM X 36MM (002-D-27036); p) D 2.7MM X 38MM (002-D-27038); q) D 2.7MM X 40MM (002-D-27040)
- Status
- Terminated
- Initiated
- 2018-04-06
- Posted
- 2018-06-07
- Terminated
- 2021-06-30
- Reason
- Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Root cause
- Device Design
- Action
- On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
- Quantity
- 935
- Distribution
- U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
- Lot, serial or model codes
- (UDI)/Lot: a) (8056099645782) B1075108; b) (18056099645799) B1075105 c) (18056099645805) B1071598, B1075178, B1075178; d) 18056099645812) B1069675, B1075431, B1075431; e) (18056099645829) B1071545, B1075435; f) (18056099645836) B1071540, B1075839; g) (18056099645843) B1068567, B1075827, B1075827; h) (18056099645850) B1082175, B1082175; i) (18056099645867) B1082237, B1068549, B1082237; j) (18056099645874) B1075060, B1082182; k) (18056099645881) B1068574, B1082236; l) (18056099645898) B1204990, B1071594, B1082232; m) (18056099645904) B1204993, B1071543, B1082222; n) (18056099645911) B1071542, B1195742, B1083645; o) (18056099645928) B1071539, B1083644, B1195708, B1083644; p) (18056099645935) B1204989, B1071541, B1089895; q) (18056099645942) B1069679, B1195752, B1069679, B1082235
- 510(k) numbers
- ["K140876"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 2183449
- Firm city
- Lewisville
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28