GE Healthcare: GE Healthcare Precision RXi 23A/32A Analog System

Recall number
Z-2121-2015
Recalling firm
GE Healthcare
Product
GE Healthcare Precision RXi 23A/32A Analog System
Status
Terminated
Initiated
2015-06-12
Posted
2015-07-20
Terminated
2017-08-03
Reason
A reported incident of a monitor that fell from the suspension. A fall of a monitor could result in an injury to a patient or operator. Recall affects the single and dual LCD monitor suspensions used on Precision 500D Classical R & F system, Precision RXi 23A/32A Analog System, Precision RXi Digital system and Precision MPi fluoroscopic imaging systems.
Root cause
Nonconforming Material/Component
Action
GE Healthcare sent an "Urgent Medical Device Correction" letter dated June 12, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Perform a visual check before and after each use for the presence of the four (4) mounting screws as shown in the pictures below. Check the secureness of these screws by hand. If these screws are present and secure, you may continue use of the system. If any of these screws are missing or can easily be turned by hand, follow the precautions below before you continue use with the monitor(s). 1. Position the monitor suspension in the most frequently used position and limit further movement as much as possible. 2. Clearly inform (e.g. through signage and verbal instructions) the users and other auxiliary personnel who may come in contact with the system to not move the monitor suspension once in position. 3. Contact your GE Healthcare representative to inform them of any loose or missing screws. For questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity
788 (701 US; 87 OUS)
Distribution
Worldwide Distribution - US Nationwide Distribution including GU and DC., and to the countries of : Australia, Bulgaria, Canada, Chile, Dominican Republic, Germany, Ireland, Italy, Kuwait, Lebanon, Portugal, Romania, Russia, Saudi Arabia, Slovakia, Spain, Taiwan,Turkey, United Kingdom, Venezuela and Yemen.
Lot, serial or model codes
Mfg Lot or Serial # 00000001386MC2 00000002109MC7 00000001010MC8 00000001337MC5 00000002212MC9 00000001720MC2 00000001098MC3 00000001454MC8 00000001082MC7 00000001266MC6 00000001656MC8 00000003661MC6
510(k) numbers
["K082243"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register