Skip to the content

FDA Medical Device Recalls (openFDA)

Burlington Medical, LLC: Kilt, Single Item, Product Number Kilt, Part Numbers ECREKILTF, ECREKILTXLF, REKILTA, REKILTB, REKILTXLF, REKILTXLM, REKILT*M, SEKILTB, SEKILTC, SEKILT*F, SEKILT*M Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.

Recall number
Z-2118-2019
Recalling firm
Burlington Medical, LLC
Product
Kilt, Single Item, Product Number Kilt, Part Numbers ECREKILTF, ECREKILTXLF, REKILTA, REKILTB, REKILTXLF, REKILTXLM, REKILT*M, SEKILTB, SEKILTC, SEKILT*F, SEKILT*M Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.
Status
Terminated
Initiated
2019-06-07
Posted
2019-07-30
Terminated
2023-04-28
Reason
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.
Root cause
Under Investigation by firm
Action
Urgent Medical Device Recall notification letters were sent to customers on 6/7/19. Please take the following actions: Cease using recalled products immediately, quarantine the products, and prepare the products for return shipment to Burlington Medical. Contact Burlington Medical at (800) 221-3466 for specific return instructions. Note that Burlington Medical Customer Care will work with you to facilitate the return of recalled products, including by issuing you a Recall Tracking number. Burlington Medical will replace all recalled products at no cost to you. The replacement product will be the same size and model as the recalled product; if the same size and model is not available or has been discontinued, then Burlington Medical will replace the recalled product with a Burlington Medical product that is similar to the recalled product. Our Customer Care team will arrange for free shipment of the replacement product. If you have any questions, please call your Burlington Medical representative or Burlington Medical Customer Care at (800) 221-3466. The recall was expanded an additional Urgent Medical Device Recall notification letters were sent to customers on 7/3/19.
Quantity
37
Distribution
US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.
Lot, serial or model codes
Serial Numbers: 387197 384398 387814 418788 419871 387879 385725 385749 385732 387887 382947 419840 418873 423988 385738 387937 444696 418901 381799 419884 379705 386112 416433 415529 443261 418973 419967 419814 423980 419787 419892 391089 391128 385744 379537 417552 414129
510(k) numbers
not on file
PMA numbers
not on file
Product code
EAJ
Firm FEI
1000120804
Firm city
Newport News
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register