Abbott Gmbh & Co. KG: ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.

Recall number
Z-2117-2020
Recalling firm
Abbott Gmbh & Co. KG
Product
ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.
Status
Open, Classified
Initiated
2020-04-24
Posted
2020-05-12
Terminated
not on file
Reason
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.
Root cause
Under Investigation by firm
Action
On April 24, 2020, Abbott issued a Product Correction letter to all customers who received shipments of the three impacted lots of ARCHITECT iGenatamicin Reagent kits. U.S. customers were sent the letter via FedEx priority overnight express. In the U.S., the customers will be requested to return to Abbott a Customer Reply Form acknowledging the receipt or a request for assistance.
Quantity
6087 Reagent Kits
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CT, DE, FL, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MT, ND, NE, NJ, NV, NY, OH, OK, PA, RI, SC, TN,TX, UT, VA, VT, WA, WI, PR. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Belarus, Belgium, Brazil, Brunei, Canada, Cayman Islands, Chile, China, Costa Rica, Croatia, Curacao, Czech Republic, Faroe Islands, France, Germany, Greece, Greenland, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Namibia, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Philippines, Poland, Portugal, Republic of Yemen, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tunisia, UAE, United Kingdom and Uruguay.
Lot, serial or model codes
MODEL: 1P31-25 LOT NUMBERS: 04281BE00, 08510BE00 and 11221BE00 Shelf life per LOT: 04281BE00;2020-MAY-21 08510BE00; 2020-AUG-08 11221BE00;2020-NOV-18
510(k) numbers
["K102699"]
PMA numbers
not on file
Product code
LCD
Firm FEI
3002809144
Firm city
Wiesbaden
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register