Medtest Holdings, Inc.: Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.
- Recall number
- Z-2116-2017
- Recalling firm
- Medtest Holdings, Inc.
- Product
- Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.
- Status
- Terminated
- Initiated
- 2012-02-15
- Posted
- 2017-05-18
- Terminated
- 2017-06-02
- Reason
- A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.
- Root cause
- Error in labeling
- Action
- Technical Bulletin Notification letter was sent 02/15/12 by email to all customers and distributors. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-757-5313.
- Quantity
- 119.168L across All lots
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Catalog numbers: HA461, HA761 , HA961 and lot codes:022204, 034801, 102401, 119503, and 134101.
- 510(k) numbers
- ["K941132"]
- PMA numbers
- not on file
- Product code
- CIT
- Firm FEI
- 1827821
- Firm city
- Canton
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28