Heartsine Technologies, Limited: Pad-Pak and Pedi-Pak accessories for the HeartSine Samaritan PAD 300P and 350P device (SAM 300P and 350P). Automated External Defibrillator electrode supplies. The Pedi-Pak is specifically designed for use with the HeartSine Samaritan PAD 300P and 350P device. It contains defibrillator electrodes in a sealed foil pouch and a battery pack in a sealed plastic housing.

Recall number
Z-2115-2014
Recalling firm
Heartsine Technologies, Limited
Product
Pad-Pak and Pedi-Pak accessories for the HeartSine Samaritan PAD 300P and 350P device (SAM 300P and 350P). Automated External Defibrillator electrode supplies. The Pedi-Pak is specifically designed for use with the HeartSine Samaritan PAD 300P and 350P device. It contains defibrillator electrodes in a sealed foil pouch and a battery pack in a sealed plastic housing.
Status
Terminated
Initiated
2014-06-12
Posted
2014-07-24
Terminated
2015-04-16
Reason
A small number of sealed foil pouches containing the electrodes were found to be difficult to open, resulting in potential delay in therapy.
Root cause
Packaging process control
Action
HeartSine notified distributors in the United States by e-mail on June 12, 2014 and by follow-up letter on June 13, 2014. Distributors were given a letter to use in communications with customers on June 17, 2014. Customers were instructed to return affected product to there distributors. Distributors were asked to return the affected product to HeartSine. In order to effect an immediate return, replacement Pad-Paks and Pedi-Paks have already been delivered to all respective distributors. A seconde letter, dated June 30th, was sent to end users to clarify where on the label to find lot specific information. For question contact Quality Manager at +44 (0)28 90 93 94 19.
Quantity
3245
Distribution
Worldwide Distribution - US (nationwide) and in Canada, Mexico and Latin America.
Lot, serial or model codes
Pad-Pak 01 & Pad-Pak 02 ¿ Adult Lot Numbers: A1785 to A1805 ¿ Pediatric Lot Numbers: P433 to P445.
510(k) numbers
["K052465","K123881"]
PMA numbers
not on file
Product code
MKJ
Firm FEI
3004123209
Firm city
Belfast
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register