Skip to the content

FDA Medical Device Recalls (openFDA)

Acumed LLC: Right Locking 1st MTP Revision Fusion Plate. Product labeled in part, "ACUMED 5885 NW Cornelius pass Road Hillsboro, OR. 97124-9432... Size: Right Locking 1st MTP Revision Fusion Plate... REF 70-0019-S... LOT W58122".

Recall number
Z-2115-2008
Recalling firm
Acumed LLC
Product
Right Locking 1st MTP Revision Fusion Plate. Product labeled in part, "ACUMED 5885 NW Cornelius pass Road Hillsboro, OR. 97124-9432... Size: Right Locking 1st MTP Revision Fusion Plate... REF 70-0019-S... LOT W58122".
Status
Terminated
Initiated
2008-02-14
Posted
2008-09-20
Terminated
2010-02-03
Reason
Product sterility may be compromised.
Root cause
Packaging process control
Action
The firm issued four separate Urgent Notification: Product Removal were issued by the firm on 2-14-2008, 2-20-2008, 2-26-2008, and 3-25-2008 for the issue of poorly sealed sterile pouches. The notifications were reportedly sent by either FAX or e-mail. The firm explained the issue and requested customers to return the specified products. On 7/16/08, the firm issued a second recall notification to their customers. This letter informed the customer of the recall, listed the products affected, stated the reason for the recall, tells customers, "return existing inventories of the recalled products to Acumed... If you further distributed this product, please notify those customers/consignees of this recall. This recall is to be extended to the user level." The letter includes an effectiveness form to be filled out by the customer and returned to Acumed.
Quantity
3 pcs
Distribution
Worldwide Distribution ---- USA including states of WA, NE, NC, UT, FL, CA, OR, TX, NY, IL, OH, MT, and Puerto Rico, and countries of United Kingdom, Australia, and Greece.
Lot, serial or model codes
Work order number/Lot: W58122 (W=work order, ###=sequential number); Expiration date: 2014-11 (Year/Month)
510(k) numbers
["K033639"]
PMA numbers
not on file
Product code
HWC
Firm FEI
1000125930
Firm city
Hillsboro
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register