FDA Medical Device Recalls (openFDA)
Philips Medical Systems (Cleveland) Inc: Ingenuity Flex, model # 728317; Intended to produce cross- sectional images of the body.
- Recall number
- Z-2113-2016
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Ingenuity Flex, model # 728317; Intended to produce cross- sectional images of the body.
- Status
- Terminated
- Initiated
- 2016-06-21
- Posted
- not on file
- Terminated
- 2018-07-02
- Reason
- Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.
- Root cause
- Nonconforming Material/Component
- Action
- Philips is informing the consignee of this issue through: Customer Information Letter (CIL) 72800656 . International consignees will be contacted by Philips Healthcare Global Market representatives, who will distribute the CIL. A Field Service Engineer (FSE) will contact the customers to schedule a time to conduct the appropriate correction through Field Change Order (FCO) 72800656 once it is available. This FCO will address the above identified issue. A response form sent to the customer via certified mail. This package will also contain a return receipt.
- Quantity
- 10 units
- Distribution
- Distributed in China, Denmark, Germany, Hungary, Italy, Philippines, Spain, Sweden, Switzerland, Thailand, and United Kingdom.
- Lot, serial or model codes
- 901475, 901476, 901495, 901505, 901509, 901513, 901529, 901532, 901539, 901541
- 510(k) numbers
- ["K033326"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28