Skip to the content

FDA Medical Device Recalls (openFDA)

Sunrise Medical (US) LLC: Padded Swing Away Arm Rest with Receiver; Uses on wheelchair models: Quickie Q7 (EIR4), GT (EIR3), GTi (EIT4) and Zippie Zone (EIRZ1). Product Usage: To empower physically challenged persons by providing a means of mobility. This includes temporary and permanent conditions.

Recall number
Z-2113-2012
Recalling firm
Sunrise Medical (US) LLC
Product
Padded Swing Away Arm Rest with Receiver; Uses on wheelchair models: Quickie Q7 (EIR4), GT (EIR3), GTi (EIT4) and Zippie Zone (EIRZ1). Product Usage: To empower physically challenged persons by providing a means of mobility. This includes temporary and permanent conditions.
Status
Terminated
Initiated
2012-06-26
Posted
2012-07-31
Terminated
2012-11-01
Reason
Reports of injury when the Padded Swing Away Armrest is used for full body weight. Owner's manual states that armrest is only designed for resting of forearm.
Root cause
Labeling design
Action
Sunrise Medical sent a Safety Notice letter dated June 25, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter request that clinicians/suppliers post this safety notice in a conspicuous location, making customers and therapists aware of the issue. The letter also encourage end users to read the owner's manual and follow the recommenations for proper usage of the recalled product. Customers are instructed to complete and return the Supplier Notification and Response Form. For questions contact Sunrise Medical Regulatory Affairs at 888-208-4901.
Quantity
14, 035
Distribution
Worldwide Distribution - US (nationwide) and the countries of Argentina, Brazil, Columbia, Costa Rica, Mexico, Panama, Uruguay and Venezuela.
Lot, serial or model codes
Receiver Product number 109207.
510(k) numbers
["K973673"]
PMA numbers
not on file
Product code
IOR
Firm FEI
2937137
Firm city
Fresno
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register