Spacelabs Healthcare, Llc: BleaseSirius Anaesthesia System, gas machine. Blease Beech House, Chiltern Court Asheridge Road, Chesham, Bucks HP5 2PX, England Intended for use in the hospital environment and operating room.
- Recall number
- Z-2113-2010
- Recalling firm
- Spacelabs Healthcare, Llc
- Product
- BleaseSirius Anaesthesia System, gas machine. Blease Beech House, Chiltern Court Asheridge Road, Chesham, Bucks HP5 2PX, England Intended for use in the hospital environment and operating room.
- Status
- Terminated
- Initiated
- 2010-06-24
- Posted
- 2010-07-29
- Terminated
- 2010-12-03
- Reason
- Potential for the touch screen on the ventilator portion of the device to go blank when touched. Possibility that this failure may also cause mechanical ventilation to stop.
- Root cause
- Other
- Action
- On 06/24/2010, Spacelabs began mailing the URGENT-MEDICAL DEVICE CORRECTION letter to their US consignees and e-mailed the same letter to their foreign consignees. The letter describes the touch screen of the BleaseSirius Anesthesia machine goes blank when touched and the possibility that this failure may cause the mechanical ventilation to stop. The consignees are asked to weigh the benefits versus the risk when deciding to use the device or not until the firm upgrades the device. The letter also instructs the consignees to notify their staff about the possibility of the display failure to prevent any adverse event that may occur. The letter states that their field service engineer will contact consignees to schedule a time to visit and upgrade all affected devices at no cost. The US consignees are advised to call 1-800-522-7025, Select 2, and the foreign consignees are advised to call 1-425-657-7200 x5089 for technical support or questions regarding the recall.
- Quantity
- 80 units total
- Distribution
- Worldwide Distribution -- USA, including states of IA, GA, LA, NC, TN, and TX and countries of AUSTRALIA, CHILE, CHINA, INDIA, ITALY, PHILIPPINES, MEXICO, POLAND, SINGAPORE, THAILAND, TURKEY, UNITED KINGDOM, and VENEZUELA.
- Lot, serial or model codes
- Serial numbers: DOMESTIC - siri-001852, siri-002292, SIRI-002081, SIRI-002082, SIRI-002085, SIRI-002089, SIRI-002091, SIRI-002097, SIRI-002099, SIRI-002101, SIRI-002102, SIRI-001851, SIRI-001818, SIRI-001819, SIRI-001821, SIRI-001822, SIRI-001823, SIRI-001824, SIRI-001825, SIRI-001826, SIRI-001827, SIRI-001828, SIRI-001829, SIRI-002106, and SIRI-002108. INTERNATIONAL - siri-001861, siri-001853, siri-001855, siri-001857, SIRI-001905, SIRI-002363, SIRI-001948, SIRI-001949, SIRI-001950, SIRI-001951, SIRI-001952, SIRI-001846, SIRI-001847, SIRI-001848, SIRI-001849, SIRI-001921, SIRI-001961, SIRI-001963, SIRI-001964, siri-002139, SIRI-002162, SIRI-002298, SIRI-002349, SIRI-002354, SIRI-001611, SIRI-001612, SIRI-001614, SIRI-001615, SIRI-001616, SIRI-001617, SIRI-002157, siri-002135, SIRI-2302, siri-001903, siri-001904, siri-001911, SIRI-002041, SIRI-001858, SIRI -002150, SIRI-002156, siri-002166, siri-002168, siri-002169, siri-002171, siri-002201, SIRI-001854, SIRI-001785, SIRI-001786, SIRI-001954, SIRI-001850, SIRI-001862, SIRI-002159, SIRI-002304, SIRI-002315, and SIRI-002224.
- 510(k) numbers
- ["K051629"]
- PMA numbers
- not on file
- Product code
- BSZ
- Firm FEI
- 3010157426
- Firm city
- Issaquah
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28