Skip to the content

FDA Medical Device Recalls (openFDA)

Spacelabs Healthcare, Llc: BleaseSirius Anaesthesia System, gas machine. Blease Beech House, Chiltern Court Asheridge Road, Chesham, Bucks HP5 2PX, England Intended for use in the hospital environment and operating room.

Recall number
Z-2113-2010
Recalling firm
Spacelabs Healthcare, Llc
Product
BleaseSirius Anaesthesia System, gas machine. Blease Beech House, Chiltern Court Asheridge Road, Chesham, Bucks HP5 2PX, England Intended for use in the hospital environment and operating room.
Status
Terminated
Initiated
2010-06-24
Posted
2010-07-29
Terminated
2010-12-03
Reason
Potential for the touch screen on the ventilator portion of the device to go blank when touched. Possibility that this failure may also cause mechanical ventilation to stop.
Root cause
Other
Action
On 06/24/2010, Spacelabs began mailing the URGENT-MEDICAL DEVICE CORRECTION letter to their US consignees and e-mailed the same letter to their foreign consignees. The letter describes the touch screen of the BleaseSirius Anesthesia machine goes blank when touched and the possibility that this failure may cause the mechanical ventilation to stop. The consignees are asked to weigh the benefits versus the risk when deciding to use the device or not until the firm upgrades the device. The letter also instructs the consignees to notify their staff about the possibility of the display failure to prevent any adverse event that may occur. The letter states that their field service engineer will contact consignees to schedule a time to visit and upgrade all affected devices at no cost. The US consignees are advised to call 1-800-522-7025, Select 2, and the foreign consignees are advised to call 1-425-657-7200 x5089 for technical support or questions regarding the recall.
Quantity
80 units total
Distribution
Worldwide Distribution -- USA, including states of IA, GA, LA, NC, TN, and TX and countries of AUSTRALIA, CHILE, CHINA, INDIA, ITALY, PHILIPPINES, MEXICO, POLAND, SINGAPORE, THAILAND, TURKEY, UNITED KINGDOM, and VENEZUELA.
Lot, serial or model codes
Serial numbers: DOMESTIC - siri-001852, siri-002292, SIRI-002081, SIRI-002082, SIRI-002085, SIRI-002089, SIRI-002091, SIRI-002097, SIRI-002099, SIRI-002101, SIRI-002102, SIRI-001851, SIRI-001818, SIRI-001819, SIRI-001821, SIRI-001822, SIRI-001823, SIRI-001824, SIRI-001825, SIRI-001826, SIRI-001827, SIRI-001828, SIRI-001829, SIRI-002106, and SIRI-002108.   INTERNATIONAL - siri-001861, siri-001853, siri-001855, siri-001857, SIRI-001905, SIRI-002363, SIRI-001948, SIRI-001949, SIRI-001950, SIRI-001951, SIRI-001952, SIRI-001846, SIRI-001847, SIRI-001848, SIRI-001849, SIRI-001921, SIRI-001961, SIRI-001963, SIRI-001964, siri-002139, SIRI-002162, SIRI-002298, SIRI-002349, SIRI-002354, SIRI-001611, SIRI-001612, SIRI-001614, SIRI-001615, SIRI-001616, SIRI-001617, SIRI-002157, siri-002135, SIRI-2302, siri-001903, siri-001904, siri-001911, SIRI-002041, SIRI-001858, SIRI -002150, SIRI-002156, siri-002166, siri-002168, siri-002169, siri-002171, siri-002201, SIRI-001854, SIRI-001785, SIRI-001786, SIRI-001954, SIRI-001850, SIRI-001862, SIRI-002159, SIRI-002304, SIRI-002315, and SIRI-002224.
510(k) numbers
["K051629"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
3010157426
Firm city
Issaquah
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register