FDA Medical Device Recalls (openFDA)
Covidien LLC: Surgical Kit Sterile-Kits containing the Devon Light Glove Catalog Numbers/Description: 573326 7497-HLS Surgical Kit 573328 7437-NRF SURGICAL ASC KIT 573343 7499-88 SURGICAL KIT with Light Gloves 573346 7499-TLG Surgical Kit with Edge 573359 7499-HLW Surgical KIT 573368 7494-CAB2 SURGICAL KIT 573776 7520-BHP SURGICAL KIT 573777 7527-BHB SURGICAL KITS 573798 7496-8KB Surgical KIT 31144507 7519 Minor Surgical Kit 31175089 7497-8T6 Surgical Kit 31321097 7697 Surgical Set Up Kit 31451480 7493-SFW SURGICAL KIT 31460432 7596-SHH Surgical KIT 50000947 7413-CAD Surgical KIT 50001029 7417-PCP Surgical KIT 50001052 7427-CFS Surgical KIT
- Recall number
- Z-2112-2015
- Recalling firm
- Covidien LLC
- Product
- Surgical Kit Sterile-Kits containing the Devon Light Glove Catalog Numbers/Description: 573326 7497-HLS Surgical Kit 573328 7437-NRF SURGICAL ASC KIT 573343 7499-88 SURGICAL KIT with Light Gloves 573346 7499-TLG Surgical Kit with Edge 573359 7499-HLW Surgical KIT 573368 7494-CAB2 SURGICAL KIT 573776 7520-BHP SURGICAL KIT 573777 7527-BHB SURGICAL KITS 573798 7496-8KB Surgical KIT 31144507 7519 Minor Surgical Kit 31175089 7497-8T6 Surgical Kit 31321097 7697 Surgical Set Up Kit 31451480 7493-SFW SURGICAL KIT 31460432 7596-SHH Surgical KIT 50000947 7413-CAD Surgical KIT 50001029 7417-PCP Surgical KIT 50001052 7427-CFS Surgical KIT
- Status
- Terminated
- Initiated
- 2015-04-16
- Posted
- 2015-07-20
- Terminated
- 2017-06-28
- Reason
- Devon Light Gloves contain splits or holes compromising the sterility
- Root cause
- Process control
- Action
- Medtronic/Covidien initiated recall on April 16, 2015 via FedEx and certified mail Customers were instructed to discontinue use of the product and return all inventory and kits to Medtronic (Covidien) -Attention: Devon" Light Glove 110 Kendall Park Lane. Atlanta, GA 30336. If you have any questions contact your Medtronic representative or Medtronic Customer Service, Monday through Friday, 8am to 6:30pm ET, at (800) 882-5878.
- Quantity
- 114,813 kits
- Distribution
- Worldwide Distribution-US (nationwide) including the Virgin Islands and the countries of Canada, Australia, New Zealand, Costa Rica, Austria, Belgium, Bulgaria, Czech Republic, Denmark, Finland , France, Germany, Netherlands, Sweden, Ireland, ISRAEL,Spain, Italy, Jordan, Kenya, Kuwait, United Kingdom, Lebanon, Liechtenstein, Luxembourg, Monaco, South Africa, Norway, Poland, Portugal, Qatar, Russia, Saudi Arabia, Slovenia, Switzerland, Turkey, and UAE.
- Lot, serial or model codes
- Lot number begins 508xxxx or lower
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FSY
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28