Skip to the content

FDA Medical Device Recalls (openFDA)

GE Healthcare, LLC: Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.

Recall number
Z-2111-2020
Recalling firm
GE Healthcare, LLC
Product
Giraffe Shuttle. Upgrade Kits with part number 2109672-001 and 2109673-001 GTIN: 00840682110808.
Status
Terminated
Initiated
2020-04-02
Posted
2020-05-22
Terminated
2024-05-06
Reason
During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Root cause
Device Design
Action
GE Healthcare notified customers on 04/02/2020 via "Urgent Medical Device Correction" letter. The recall letter identified the affected product and serial numbers. The firm instructed the users to discontinue use of the product if at any time they smell a gas odor. GE Healthcare will correct all affected products at a later date when the correction is available at no cost.
Quantity
204 units.
Distribution
US Nationwide distribution in the states of AZ, CA, CO, DE, FL, GA, ID, IL, IN, LA, MD, MI, MS, NC, ND, NM, NV, NY, OH, Ok, OR, PA, SC, TN, TX, VA, WA, WI and WV.
Lot, serial or model codes
Serial numbers: GSHW60086, GSHX60001, GSHX60014, GSHY70002-GSHY70195, GSHZ70101-GSHZ70108, MAAP50034, MAAQ50170, MAAR50021, MAAT50013, MAAT60045, MAAU61029, MAAU61040 and MAAV6059.
510(k) numbers
["K101788"]
PMA numbers
not on file
Product code
FMZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register