Philips Medical Systems (Cleveland) Inc: Brilliance CT Big Bore Oncology, model #728243; Intended to produce cross- sectional images of the body.

Recall number
Z-2111-2016
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance CT Big Bore Oncology, model #728243; Intended to produce cross- sectional images of the body.
Status
Terminated
Initiated
2016-06-21
Posted
not on file
Terminated
2018-07-02
Reason
Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack and cause injury to patients, operators, by standers or service personnel.
Root cause
Nonconforming Material/Component
Action
Philips is informing the consignee of this issue through: Customer Information Letter (CIL) 72800656 . International consignees will be contacted by Philips Healthcare Global Market representatives, who will distribute the CIL. A Field Service Engineer (FSE) will contact the customers to schedule a time to conduct the appropriate correction through Field Change Order (FCO) 72800656 once it is available. This FCO will address the above identified issue. A response form sent to the customer via certified mail. This package will also contain a return receipt.
Quantity
3 units
Distribution
Distributed in China, Denmark, Germany, Hungary, Italy, Philippines, Spain, Sweden, Switzerland, Thailand, and United Kingdom.
Lot, serial or model codes
921216
510(k) numbers
["K033357"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register