Skip to the content

FDA Medical Device Recalls (openFDA)

I-Flow Corporation: I-Flow Corporation Easy Pump LT 60-24 (60 ml x 2 ml/hr), REF: 04434366, assembled in Mexico. The Easypump LT is indicated for continuous and/or intermittent infusion of medication for general infusion use, including antibiotic delivery, chemotherapy and pain management.

Recall number
Z-2111-2009
Recalling firm
I-Flow Corporation
Product
I-Flow Corporation Easy Pump LT 60-24 (60 ml x 2 ml/hr), REF: 04434366, assembled in Mexico. The Easypump LT is indicated for continuous and/or intermittent infusion of medication for general infusion use, including antibiotic delivery, chemotherapy and pain management.
Status
Terminated
Initiated
2008-02-11
Posted
2009-09-14
Terminated
2010-02-04
Reason
I-Flow Corporation received two complaints involving Easypump LT 60-24 (Lot 612996) containing a 100 ml x 2 ml/hr pump in the sterile package instead of a 60 ml x 2 ml/hr pump (Illustration was included to show consignee what both the correct and incorrect part looks like). Lot 612996 was manufactured on January 6, 2006. As of February 7, 2008, there have been no further complaints received on t
Root cause
Labeling mix-ups
Action
I-Flow Corp. issued an "Important Recall Information" letter dated February 7, 2008 informing Consignees to quarantine and return any remaining inventory and notify any end users of the affected lot. For further information, contact I-Flow Corp. at 1-949-206-2653.
Quantity
4,450 Distributed
Distribution
Nationwide Distribution to B. Braun, Spain, who then further distributed to Italy, Germany, Netherland, Latvia, Spain, Portugal and Singapore.
Lot, serial or model codes
Lot Number: 612996.
510(k) numbers
["K052117"]
PMA numbers
not on file
Product code
MEB
Firm FEI
3011270181
Firm city
Lake Forest
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register