Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott: Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac).

Recall number
Z-2110-2023
Recalling firm
Abbott
Product
Abbott Amplatzer Steerable Delivery Sheath (ASDS), REF ASDS-14F-075, is a percutaneous catheter (cardiac).
Status
Open, Classified
Initiated
2023-06-12
Posted
2023-07-11
Terminated
not on file
Reason
There is an overall reported incidence rate of 0.77% related to observed or potential air embolism during procedures in which the ASDS product was used.
Root cause
Device Design
Action
Abbott issued a MEDICAL DEVICE RECALL notice to its consignees beginning on 06/12/2023. The notice explained the problem with the device, the risk, and requested the return of all units. Should you have questions about this issue, please contact your local Abbott Representative or Abbott Support at 1-800-544-1664 (Option 2) (U.S.), 7:00 a.m. - 7:00 p.m. Central Time, Monday through Friday.
Quantity
675 units
Distribution
Worldwide distribution - US Nationwide and the countries of Slovakia, Czech Republic.
Lot, serial or model codes
UDI/DI 05415067036025, Batch Numbers: 8597228, 8598983, 8601676, 8611443, 8619197, 8627877, 8650955, 8663679, 8795668, 8815423, 8824607, 8830530, 8846835
510(k) numbers
["K220340"]
PMA numbers
not on file
Product code
DQY
Firm FEI
3005334138
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register