Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corporation: Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.

Recall number
Z-2110-2021
Recalling firm
Baxter Healthcare Corporation
Product
Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Status
Terminated
Initiated
2021-06-04
Posted
2021-08-06
Terminated
2024-09-25
Reason
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Root cause
Software design
Action
Baxter initiated the recall by letter on 06/04/2021. The letter described the problem and the hazard involved, and requested the consignee Baxter Technical Assistance in advance of making changes to your network or server infrastructure supporting the pumps or the Baxter Gateway, as indicated in Baxter's Gateway Server Installation Handoff document which customers receive upon installation. The firm is developing a software patch for all Spectrum IQ infusion pumps.
Quantity
130135 units
Distribution
United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
Lot, serial or model codes
All Serial Numbers
510(k) numbers
["K173084"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register