The Binding Site Group, Ltd.: Optilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)

Recall number
Z-2110-2020
Recalling firm
The Binding Site Group, Ltd.
Product
Optilite Optilite Rheumatoid Factor Kit, Ref: LK151.OPT.A, IVD, Rx Only, UDI: 05051700020015 (kit ea. containing four components)
Status
Terminated
Initiated
2019-10-07
Posted
not on file
Terminated
2022-01-11
Reason
Assigned Rheumatoid Factor (RF) calibrator value for calibrator component included in kit used to aid in the diagnosis of rheumatoid arthritis over recovers by 7% in comparison to the reference material RF serum. This could result in reporting abnormal (false positive) results, which may lead to unnecessary additional serological testing and a delay in treatment.
Root cause
Process control
Action
The field safety notice was issued from the firm's UK office to their US office. The US office notified affected end users in the US via Field Safety Notice, dated 09 October 2019 that was sent via email on 9 Oct 2019. The FSN stated the following: 1) That it has been identified that the assigned Rheumatoid Factor (RF) calibrator value, for LK151.0PT.A lot 428224, over recovers by 7% in comparison to the reference material RF serum, NIBSC. Ref 64/002 (WHO Standard 64/2). 2) The firm has assessed risk of this over recovery and determined that there is no clinical risk, and that the kit can continue to be used providing the control results are within their specified acceptance ranges. There is however, a risk of reporting abnormal results which may lead to unnecessary additional serological testing. 3) Although the affected kit, lot 428224, may continue to be used, The Binding Site has calculated an adjustment factor of 0.93 that may be applied to sample results in order to align results to the reference material. They advise that to apply the adjustment factor, the final result for the sample should be multiplied by 0.93. Calibrator and control values for future lots of LK151.0PT.A, will be aligned with the reference material. 4) An e-back form (TSWS18) was provided to affected customers along with the FSN. The FSN requests the customer to complete, sign and return the e-back form to Technical.Support@bindingsite.com or to their local Binding Site representative within one week of receiving the notification to confirm that they have received and understood the FSN. 5) If further information is required, contact local Binding Site Representative or Technical Support Group, UK: +44(0) 1214569696, Technical.support@bindingsite.com
Quantity
478 kits
Distribution
US: AL, AR, AZ, FL, GA, KY, MA, MD, MI, MN, MS, NC, NE, NY, OR, PA, TX, VT, WA, WI OUS: None
Lot, serial or model codes
Kit Lot Number: 428224
510(k) numbers
["K162263"]
PMA numbers
not on file
Product code
DHR
Firm FEI
3002808340
Firm city
Birmingham
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register