Skip to the content

FDA Medical Device Recalls (openFDA)

Alere North Americal: Clearview iFOB, Clearview - Clearview iFOB Specimen Collection (accessory) Product Part Number: 8116KCV Intended Use: Immunochemical device intended for the qualitative detection if fecal occult blood in human feces by laboratories or physician offices

Recall number
Z-2110-2011
Recalling firm
Alere North Americal
Product
Clearview iFOB, Clearview - Clearview iFOB Specimen Collection (accessory) Product Part Number: 8116KCV Intended Use: Immunochemical device intended for the qualitative detection if fecal occult blood in human feces by laboratories or physician offices
Status
Terminated
Initiated
2011-03-21
Posted
2011-04-28
Terminated
2013-03-07
Reason
Potential for false negatives test results and negative/faint positive controls
Root cause
Other
Action
Alere sent a Voluntary Product Recall letter dated March 21, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to indicate zero on the attached Recall Reply Form, sign, and fax to 1-877-877-0147, if they have no remaining product. For any remaining product customers were instructed to discontinue use and indicate the quantity of kits in the attached Recall Reply Form so that the firm can arrange for them to receive credit. Customers were instructed to destroy any affected units at their site. For any further distribution of product, customers will need to be contacted. Physicians and hospitals were advised to retest patients with negative results. Questions relating to this Voluntary Product Recall call 1-877-597-9586.
Quantity
1800 units
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot Numbers:  0626510 0805110
510(k) numbers
["K063693"]
PMA numbers
not on file
Product code
KHE
Firm FEI
3006984151
Firm city
Orlando
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register