Skip to the content

FDA Medical Device Recalls (openFDA)

International Technidyne Corp.: RQCHRT HRT Control Level 1 & 2; Whole Blood controls for use with Hemochron Kaolin and Celite HRT Test Tubes. 20 vials Whole Blood - 2.0mL; dried 20 vials Diluent - 4.0 mL The Hemochron HRT Whole Blood Quality Controls are intended to be used to perform quality control assays using the Hemochron HRT test tubes.

Recall number
Z-2110-2010
Recalling firm
International Technidyne Corp.
Product
RQCHRT HRT Control Level 1 & 2; Whole Blood controls for use with Hemochron Kaolin and Celite HRT Test Tubes. 20 vials Whole Blood - 2.0mL; dried 20 vials Diluent - 4.0 mL The Hemochron HRT Whole Blood Quality Controls are intended to be used to perform quality control assays using the Hemochron HRT test tubes.
Status
Terminated
Initiated
2010-06-24
Posted
2010-07-28
Terminated
2010-09-23
Reason
An ingredient in the RQCHRT Whole Blood Controls for use with Hemochron Kaolin and Celite HRT Test tubes may become less potent over time and result in excessive High LQC Failures.
Root cause
Other
Action
ITC sent "URGENT MEDICAL DEVICE RECALL" letters dated June 24, 2010, via Federal Express to Laboratory Director, Clinical Laboratory Manager, or Point of Care Coordinator consignees/customers. The letter described the product, problem and action to be taken by customers. The customers were instructed to check their inventory for the product, if they have boxes or loose vials to please stop using the product, remove it from inventory and contact ITC Technical Support to arrange return. They were also instructed to complete the attached Customer Account Tracking Form, whether or not they have remaining inventory, and return it by fax, email or mail. If you have any questions, please call ITC Technical Support at 800-631-5945 (US) or 732-548-5700 (international), Extension 4707 or email techsupport@itcmed.com.
Quantity
38 in US, 3 Out of US
Distribution
Worldwide distribution: USA including states of CA, NC, TX, FL, MD, OK, NY, and NJ and country of Italy.
Lot, serial or model codes
Test kit lot numbers: H9KHR002, Exp 05/2010 and BOKHR001, Exp 08/2010; RAF 10-016
510(k) numbers
["K904003"]
PMA numbers
not on file
Product code
KFF
Firm FEI
1000526865
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28