Skip to the content

FDA Medical Device Recalls (openFDA)

Toshiba American Medical Systems Inc: Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F

Recall number
Z-2109-2017
Recalling firm
Toshiba American Medical Systems Inc
Product
Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F
Status
Terminated
Initiated
2017-02-01
Posted
not on file
Terminated
2018-06-06
Reason
It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on.
Root cause
Radiation Control for Health and Safety Act
Action
Toshiba Medical planned action to be taken to address the defect: CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Any future customer notification letters to be sent must comply with the requirements of 21 CFR 1003.21. This includes, but is not limited to, the prescribed packaging and the statement that you will remedy the defect or bring the product into compliance at no charge. 3. The corrections are made free of charge to customers. 4. You complete this recall by September 30, 2017. Report any failures regarding this recall immediately to the District Recall Coordinator. For further questions, please call (800) 521-1968
Quantity
US - 70 systems
Distribution
US Distribution to the states of :LA, IL, NY, FL, MA, TN, MO, GA, CA, DE, MO, AL, ID, MN, TX, CT, PA, MS, OH, NC, NV, W VA, WA, AR, NJ, MD, OK, IN and NH.
Lot, serial or model codes
S/N : 8181622393
510(k) numbers
["K152785"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2020563
Firm city
Tustin
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register