Philips Ultrasound, Inc.: QLAB 10 Core Module PN 453561704771. QLAB Quantification Software is a software application package designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.

Recall number
Z-2109-2014
Recalling firm
Philips Ultrasound, Inc.
Product
QLAB 10 Core Module PN 453561704771. QLAB Quantification Software is a software application package designed to view and quantify image data acquired on Philips Medical Systems ultrasound products.
Status
Terminated
Initiated
2014-07-09
Posted
2014-07-23
Terminated
2015-11-06
Reason
The QLAB a2DQ and aCMQ features have a defect that may use incorrect measurement values under specific conditions.
Root cause
Software change control
Action
Philips Medical Systems notified customers of the voluntary recall by sending a certified URGENT-Medical Device Correction letter, dated 2014 JUL 09, to all US and Canadian consignees. Philips representatives will distribute the letter internationally. Follow-up letters to wholesale customers will be provided and wholesalers are requested to forward copies of recall letters and response forms to their customers. Philips Medical Systems will also have Customer Reply Cards returned for tracking purposes. Customers are directed to call local Philips representative or Philips Customer Service at 1-866-767-7822.
Quantity
25 units total (2 units in the US and 23 units outside the US)
Distribution
Distributed in the states of CA, IA, KY, MD, NY, PA, UT, VA, WA and WI. Also distributed in Australia, Belgium, Canada, India, Italy, Japan, South Korea, Netherlands, New Zealand, Spain, Sweden, and Switzerland.
Lot, serial or model codes
*****SYSTEM BASED ID OUTSIDE THE US**********  120F-P213FA35-V41E, 279E-Q21384E8-P41D, 3FBF-X2649F25-J4BF,  47FD-Q1FD112B-J3F0, 4A8F-U24EA0B5-R493, 5570-Q20D1AB2-W410,  5D8C-G1FC9BDE-N3EF, 684B-V2122501-R41A, 6DA2-G232AD5C-S45B,  731B-Q1FC2691-V3EE, 808E-P20C3018-I40E, 88AA-T1FBB144-W3ED,  A0CB-I24CCB81-M48F, A0ED-H1F5BC5B-P3E1, AB9B-S236CD11-W463,  BE98-V1E4C83A-P3BF, C6A3-H1FFD0F9-X3F5, CBE9-M24BE0E7-Y48D,  EEFC-Y22FED8E-S455, EF0D-I20465FB-I3FE, F475-U1F966C3-X3E8, and  V49C-R203F0AE-I3FE.  ***********Two units shipped in US and one unit shipped outside the US to distributors do not have serial numbers. There are licensing agreements with these distributors that allows the distributors to create their own licensing and serial numbers and/or tracking information.
510(k) numbers
["K152165"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3019216
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register