Skip to the content

FDA Medical Device Recalls (openFDA)

MicroAire Surgical Instruments, LLC: MicroAire SmartRelease Endoscopic System, Endoscopic Carpal Tunnel Release, SmartRelease Handpiece, containing IFU rev. D The MicroAire Carpal Tunnel Release System is for use in patients diagnosed with carpal tunnel syndrome that is not associated with, or secondary to, any other known pathology (i.e., idiopathic carpal tunnel syndrome). Preoperative x-rays of the wrist, including a carpal tunnel view, are recommended to aid in the diagnosis of associated pathology (i.e., calcific tendonitis, fracture of the hook of the hamate). This device is indicated for releasing the transverse carpal ligament (flexor retinaculum).

Recall number
Z-2107-2018
Recalling firm
MicroAire Surgical Instruments, LLC
Product
MicroAire SmartRelease Endoscopic System, Endoscopic Carpal Tunnel Release, SmartRelease Handpiece, containing IFU rev. D The MicroAire Carpal Tunnel Release System is for use in patients diagnosed with carpal tunnel syndrome that is not associated with, or secondary to, any other known pathology (i.e., idiopathic carpal tunnel syndrome). Preoperative x-rays of the wrist, including a carpal tunnel view, are recommended to aid in the diagnosis of associated pathology (i.e., calcific tendonitis, fracture of the hook of the hamate). This device is indicated for releasing the transverse carpal ligament (flexor retinaculum).
Status
Terminated
Initiated
2018-04-19
Posted
2018-05-25
Terminated
2020-04-10
Reason
The Instructions for Use (IFU) is incorrect with regards to sterilization parameters and must be replaced.
Root cause
Other
Action
Urgent Medical Device Recall letters were distributed to customers. The letters instructed customers to do the following: If you have some quantity of IFUs remaining, we're asking you to destroy your current copies, IM-SMARTRELEASE Rev. A-D, and replace them by going on-line at MicroAire at www.microaire.com/resources. Click Carpal Tunnel, then SmartRelease Instruction Manual to obtain the latest version, IM-SMARTRELEASE Rev. E. Please complete and send back page two of this letter to notification@microaire.com or via fax at 800-648-4309. If you have no quantity of IFUs remaining, there are no further activities needed by you except for the mandatory returning of page two of this letter to notification@microaire.com or via fax at 800-648-4309. For further questions, please call (800) 722-0822.
Quantity
2009
Distribution
Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of: Australia, Belgium, Brazil, Canada, Chile, Czech Republic, Denmark, France, Germany, Hong Kong, India, Iran, Ireland, Israel, Italy, Jordan, Korea, Lebanon, Mexico, Nicaragua, Poland, Russia, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, and UK.
Lot, serial or model codes
All products distributed from 1/1/13 to 2/23/18
510(k) numbers
["K912871"]
PMA numbers
not on file
Product code
HRX
Firm FEI
2020601
Firm city
Charlottesville
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register