Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: Fountain Infusion System, 4F, REF IS4-90-50, Sterile, Merit Medical Systems, Inc., South Jordan, Utah 84095. Continuous flush catheter intended to administer infusions of various therapeutic solutions into the peripheral vasculature of a patient.

Recall number
Z-2107-2008
Recalling firm
Merit Medical Systems, Inc.
Product
Fountain Infusion System, 4F, REF IS4-90-50, Sterile, Merit Medical Systems, Inc., South Jordan, Utah 84095. Continuous flush catheter intended to administer infusions of various therapeutic solutions into the peripheral vasculature of a patient.
Status
Terminated
Initiated
2008-05-14
Posted
2008-09-11
Terminated
2010-03-08
Reason
Catheters were packaged with occluding wires that were too long for the catheter. Potential for wire to not occlude the catheter appropriately, allowing the wire to exit the tip of the catheter further than intended, posing a potential risk of physical damage to the patient.
Root cause
Packaging process control
Action
Merit's six sales reps for the six hospital consignees were notified by letter on 05/14/2008. They were instructed to contact the hospitals and have them immediately quarantine any affected product for return to Merit. If you have any questions please contact Jackie Preece, Customer Service at 1-201-253-1600.
Quantity
15 units
Distribution
Class II Recall - Nationwide Distribution --- including states of MA, NV, NY, PA, and SC.
Lot, serial or model codes
Lot numbers: F496042, and F587928.
510(k) numbers
["K992231"]
PMA numbers
not on file
Product code
KRA
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register