Skip to the content

FDA Medical Device Recalls (openFDA)

Datascope Corporation: Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-TOP 5909 14 Pack QPID, Part Number 701064992

Recall number
Z-2106-2018
Recalling firm
Datascope Corporation
Product
Maquet Cardiopulmonary Bypass Custom Tubing kit BEQ-TOP 5909 14 Pack QPID, Part Number 701064992
Status
Terminated
Initiated
2016-03-11
Posted
not on file
Terminated
2019-06-07
Reason
There is a potential problem with the cuvette where failure of the cuvette would cause an error message on the monitor display of H/S DISCONNECT AT CUVETTE.
Root cause
Under Investigation by firm
Action
Urgent Safety Alert notifications were sent to customers on 3/11/16. The notices included the following instructions for customers: Please examine your inventory immediately to determine if you have the lot/batch numbers listed. If so, please following the recommended actions outlined in the attached Terumo Cardiovascular Systems Urgent Safety Alert. URGENT: Please complete attached Customer Response Form on page 3, to acknowledge that you have received the Terumo Cardiovascular Systems Urgent Safety Alert. Please fax the completed Customer Response Form to 1-973-396-3607 or send via email to fieldactions@maquet.com.
Quantity
19 total kits
Distribution
The products were distributed to the following US states: AL, OR, and TX.
Lot, serial or model codes
Batch Number 3000018640, UDI Code 10607567205373;  Batch Number 3000020516, UDI Code 10607567205373
510(k) numbers
["K080592"]
PMA numbers
not on file
Product code
DWE
Firm FEI
1000222374
Firm city
Fairfield
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register