Skip to the content

FDA Medical Device Recalls (openFDA)

ArthroCare Corporation: EVac 70 XTRA with Integrated Cable and PROcise EZ View with Integrated Cable Product Usage: PLASMA WAND devices are indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: adenoidectomy, cysts, head, neck, oral, and sinus surgery, mastoidectomy, myringotomy with effective hemorrhage control, nasal airway obstruction be reduction of hypertrophic nasal turbinates, nasopharyngeal/laryngeal indications including tracheal procedures, laryngeal polypectomy, and laryngeal lesion debulking, neck mass, papilloma keloids, submucosal palatal shrinkage, submucosal tissue shrinkage, tonsillectomy (including palatine tonsils), traditional uvulopalatoplasty (RAUP), tumors, and tissue in the uvula/soft palate for the treatment of snoring. It is intended for procedures using a conductive media, such as normal saline or Ringers lactate.

Recall number
Z-2105-2015
Recalling firm
ArthroCare Corporation
Product
EVac 70 XTRA with Integrated Cable and PROcise EZ View with Integrated Cable Product Usage: PLASMA WAND devices are indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery including: adenoidectomy, cysts, head, neck, oral, and sinus surgery, mastoidectomy, myringotomy with effective hemorrhage control, nasal airway obstruction be reduction of hypertrophic nasal turbinates, nasopharyngeal/laryngeal indications including tracheal procedures, laryngeal polypectomy, and laryngeal lesion debulking, neck mass, papilloma keloids, submucosal palatal shrinkage, submucosal tissue shrinkage, tonsillectomy (including palatine tonsils), traditional uvulopalatoplasty (RAUP), tumors, and tissue in the uvula/soft palate for the treatment of snoring. It is intended for procedures using a conductive media, such as normal saline or Ringers lactate.
Status
Terminated
Initiated
2015-06-02
Posted
2015-07-18
Terminated
2016-06-27
Reason
Wands inspected with the adaptor fixture were released to the field not meeting the dielectric (Hipot) specifications.
Root cause
Nonconforming Material/Component
Action
Smith&Nephew sent an Urgent Notification letter dated June 10, 2015 via mail/courier service with delivery tracking/confirmation to affected customers. A response form is included in the notification letter. A follow up letter will be sent to non-responding consignees. Customers were instructed to complete the acknowledgement form and return via fax or email.
Quantity
733 devices
Distribution
Worldwide Distribution - US Nationwide and the countries of Australia, Belgium, Germany, Spain, United Kingdom, Ireland, India, Italy and Singapore.
Lot, serial or model codes
Lot numbers 1093290 and 1091254
510(k) numbers
["K070374","K142999"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3010813691
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register