FDA Medical Device Recalls (openFDA)
Intuitive Surgical, Inc.: KIT,DISPOSABLE ACCESSORY,4ARM,IS3000,5 PACK; Product Usage: The intended use of the Instrument arm drape is to protect equipment from contamination and to maintain a sterile field during surgical procedures using the da Vinci Surgical Systems.
- Recall number
- Z-2105-2012
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- KIT,DISPOSABLE ACCESSORY,4ARM,IS3000,5 PACK; Product Usage: The intended use of the Instrument arm drape is to protect equipment from contamination and to maintain a sterile field during surgical procedures using the da Vinci Surgical Systems.
- Status
- Terminated
- Initiated
- 2012-07-13
- Posted
- 2012-07-27
- Terminated
- 2013-06-12
- Reason
- Specific lots of the Instrument Arm Drapes were manufactured with a sterile adaptor that may have difficulty engaging an instrument.
- Root cause
- Nonconforming Material/Component
- Action
- Intuitive Surgical sent an Urgent Medical Device Correction letters dated July 16, 2012 via Fed Ex to all affected customers in the US. The customers outside of the US were notified starting July 17, 2012. The letter identified the affected products, problem and actions to be taken. Customers were instructed to Identify affected lots, order replacement products and return both full and partial boxes of the affected products for credit as soon as unaffected products are available. All staff should be made aware of the issue. Customers were instructed to complete and return the acknowledgement form immediately. Questions and concerns should be directed to Intuitive Customer Service at 1-800-876-1310. EXPANSION: Intuitive discovered an additional lot (Lot DA121075) of Instrument Arm Drape IS2000, 20 pack that should have been included in the recall and expanded the recall on September 4, 2012. Letters to customers were sent by Fed Ex on September 5, 2012.
- Quantity
- 92390 total instrument Drapes
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Australia, Spain, India, Italy and Saudi Arabia.
- Lot, serial or model codes
- Part number: 420291-03 Lot numbers: D120795 ,D120795A , D120805, D120815, D120875, D120885, D120885A, D120935, D120935A,D120945 D121015 D121025 D121025A D121105 D121105A D121355A
- 510(k) numbers
- ["K050369"]
- PMA numbers
- not on file
- Product code
- NAY
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28