Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: VersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 Device approved and commercialized in Japan only.

Recall number
Z-2104-2025
Recalling firm
Boston Scientific Corporation
Product
VersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 Device approved and commercialized in Japan only.
Status
Open, Classified
Initiated
2025-06-10
Posted
2025-07-10
Terminated
not on file
Reason
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Root cause
Packaging change control
Action
On June 10, 2025, the firm began notifying consignees of the field action via Urgent Medical Device Removal letters. Customers are instructed to not use affected product. They should remove affected units from their inventory and segregate the units in a secure place until they can be returned to Boston Scientific.
Quantity
14,724
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
Lot, serial or model codes
UDI-DI 00685447006121 Lot 35935598
510(k) numbers
not on file
PMA numbers
not on file
Product code
DXF
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register