FDA Medical Device Recalls (openFDA)
Biomet, Inc.: Biomet Interlok 75 MM fixed I-Beam Tibial Plate with locking bar CO-CR-MO/TI 6AL 4V ALLOY, REF 141224, Sterile, BIOMET ORTHOPEDICS, WARSAW. IN I-Beam Tibia Plates are knee joint replacement prostheses intended for application with or without bone cement.
- Recall number
- Z-2104-2011
- Recalling firm
- Biomet, Inc.
- Product
- Biomet Interlok 75 MM fixed I-Beam Tibial Plate with locking bar CO-CR-MO/TI 6AL 4V ALLOY, REF 141224, Sterile, BIOMET ORTHOPEDICS, WARSAW. IN I-Beam Tibia Plates are knee joint replacement prostheses intended for application with or without bone cement.
- Status
- Terminated
- Initiated
- 2011-03-03
- Posted
- 2011-04-28
- Terminated
- 2011-07-27
- Reason
- The firm is recalling the product due to the product's packaging being labeled "I-Beam 75mm. Tibial Plate", but the package actually contains the "Interlok 67mm. CR Tibial Plate".
- Root cause
- Packaging process control
- Action
- Biomet Orthopedics Inc initiated an electronic communication with Direct Accounts on March 3, 2011 and instructed them to quarantine and return all remaining/unused devices to the firm. For questions regarding this recall call 574-372-3983.
- Quantity
- 8
- Distribution
- Worldwide Distribution -USA including FL, IN, and NC and the countries of Japan and the Netherlands
- Lot, serial or model codes
- REF 141224, LOT J100921070
- 510(k) numbers
- ["K915132"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1825034
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28