Baxter Healthcare Corporation: SIGMA Spectrum Infusion Pump, Product Code 35700BAX2

Recall number
Z-2103-2023
Recalling firm
Baxter Healthcare Corporation
Product
SIGMA Spectrum Infusion Pump, Product Code 35700BAX2
Status
Open, Classified
Initiated
2023-06-15
Posted
2023-07-13
Terminated
not on file
Reason
A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.
Root cause
Under Investigation by firm
Action
Customers were contacted via phone and email beginning on June 2, 2023. An Urgent Medical Device communication was sent to all affected customers on June 15, 2023, via U.S.P.S., first class mail. The notice explained the problem, the risk involved, and requested the following: "Actions to be taken by Customers: 1. Operators may continue to use Spectrum V8 and Spectrum IQ pumps by following on-screen instructions, or the infusion setup instructions in the Preparing the Pump and IV Sets and Programming the Pump sections and upstream occlusion alarm troubleshooting in the Alarms section of the Operator s Manual. An electronic copy of the Operator s Manual can be accessed at https://service.baxter.com/tsportal/. Please note that false upstream occlusion alarms can occur at a higher rate until the software reversion is completed. If you are unable to resolve an upstream occlusion alarm, unload and reload the set. 2. A Baxter representative will contact your facility to determine the correction plan and schedule the software reversion. The representative will work with you to determine a list of affected serial numbers at your facility. Please note you will be receiving this software reversion from Baxter at no charge." "Further information and support: If you have additional questions or experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-356-3454 (choose option 3) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time."
Quantity
3306 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
GTIN 00085412498683, Software version v8.01.01, Serial Numbers
510(k) numbers
["K230022"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register