Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES INC: 16 mm Distraction Pin, Titanium, Qty: 1 per pack; STERILE; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA Distraction Pin is designed for Anterior Cervical Fusion Procedures. Use with vertebral body distraction instruments. It is intended as a temporary fixation screw utilized during cervical spine procedures.

Recall number
Z-2103-2017
Recalling firm
MEDLINE INDUSTRIES INC
Product
16 mm Distraction Pin, Titanium, Qty: 1 per pack; STERILE; Manufactured for: Medline Industries, Inc., Mundelein, IL 60060 USA Distraction Pin is designed for Anterior Cervical Fusion Procedures. Use with vertebral body distraction instruments. It is intended as a temporary fixation screw utilized during cervical spine procedures.
Status
Terminated
Initiated
2016-12-14
Posted
not on file
Terminated
2019-07-09
Reason
Product's non-conformity involves the integrity of the seal in the sterile packaging. It is possible that the seal on the sterile packaging has been compromised resulting in a loss of sterility of the medical device contained within.
Root cause
Process control
Action
Medline Industries sent an Immediate Action Required letter dated December 14, 2016, to all affected customers with response forms via US mail, notifying them of the recall. Customers were instructed to quarantine affected product and return it to the firm. The product will be repackaged and sterilized. Customers with questions were instructed to call 866-359-1704. For questions regarding this recall call 847-643-3245.
Quantity
198 individual packs
Distribution
Nationwide Distribution
Lot, serial or model codes
Item# MDS9091616T; Lot #132638
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register