Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): Medtronic Reveal LINQ LNQ11 / PA96000

Recall number
Z-2102-2021
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Product
Medtronic Reveal LINQ LNQ11 / PA96000
Status
Open, Classified
Initiated
2021-06-01
Posted
not on file
Terminated
not on file
Reason
Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.
Root cause
Software design
Action
The firm began notifying their consignees on 06/02/2021 by letter. The letter explained the issue, provided patient management recommendations, and reported that a software update was anticipated in late 2021. For questions, contact your local Medtronic Representative or Medtronic Technical Services at 800-929-4043.
Quantity
666210 devices
Distribution
Worldwide Distribution
Lot, serial or model codes
All serial numbers distributed on or after February 23, 2017    GTIN: 643169845763, 643169845770, 643169845787, 643169845794, 643169845800, 643169845817, 643169845824, 00643169845831, 00643169845848, 00643169845855, 00643169845862, 00643169845879, 00643169845893, 00763000188382, 00763000188399, 00763000188405, 00763000188412, 00763000188429, 00763000188436, 00763000188443, 00763000188450, 00763000188467, 643169845688, 643169845695, 643169845701, 643169845718, 643169845732, 643169845749, 643169845756
510(k) numbers
["K162855"]
PMA numbers
not on file
Product code
MXD
Firm FEI
2182208
Firm city
Mounds View
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register