FDA Medical Device Recalls (openFDA)

DATEX--OHMEDA, INC.: Giraffe OmniBed, with the following Model Numbers: 6650-0004-901, 6650-0001-901, 6650-0003-901, 6650-0002-901, 6650-0094-901, 6650-0001-9901, 6650-0029-901, 6650-0080-901, 6650-0032-901, 6650-0077-901, 6650-001-901, 6650-0006-901, 6650-0129-901, 6650-0063-902, 6650-0130-901, 6650-0082-902, 6650-0061-902, 6650-0131-902, 6650-0034-902, 6650-0104-902, 6650-0066-902, 6650-0042-902, 6650-0143-901, 6650-0148-902, 6650-0152-901, 6650-0107-902, 6650-0136-902, 6650-0060-902, 6650-0056-902, 6650-0043-902. Labeled as OMNIBED ENG EBASE S-HUMID, GIRAFFE OMNIBED, GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, OMNIBED SW EBASE S-HUMID, Giraffe OmniBed Legacy, OMNIBED FR EBASE S-HUMID, OMNIBED GR EBASE S-HUMID, OMNIBED SP EBASE S-HUMID, ASSY-LAU, OMNIBED ENG EBASE S-HUMI D, Finished good - Make, OBSOLETE-OMNIBED UK EBASE S-HUMID, OMNIBED UK EBASE S-HUMID. Incubator, Neonatal

Recall number
Z-2101-2023
Recalling firm
DATEX--OHMEDA, INC.
Product
Giraffe OmniBed, with the following Model Numbers: 6650-0004-901, 6650-0001-901, 6650-0003-901, 6650-0002-901, 6650-0094-901, 6650-0001-9901, 6650-0029-901, 6650-0080-901, 6650-0032-901, 6650-0077-901, 6650-001-901, 6650-0006-901, 6650-0129-901, 6650-0063-902, 6650-0130-901, 6650-0082-902, 6650-0061-902, 6650-0131-902, 6650-0034-902, 6650-0104-902, 6650-0066-902, 6650-0042-902, 6650-0143-901, 6650-0148-902, 6650-0152-901, 6650-0107-902, 6650-0136-902, 6650-0060-902, 6650-0056-902, 6650-0043-902. Labeled as OMNIBED ENG EBASE S-HUMID, GIRAFFE OMNIBED, GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, OMNIBED SW EBASE S-HUMID, Giraffe OmniBed Legacy, OMNIBED FR EBASE S-HUMID, OMNIBED GR EBASE S-HUMID, OMNIBED SP EBASE S-HUMID, ASSY-LAU, OMNIBED ENG EBASE S-HUMI D, Finished good - Make, OBSOLETE-OMNIBED UK EBASE S-HUMID, OMNIBED UK EBASE S-HUMID. Incubator, Neonatal
Status
Open, Classified
Initiated
2023-05-19
Posted
2023-07-06
Terminated
not on file
Reason
GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices
Root cause
Nonconforming Material/Component
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following: Actions to be taken by Customer/User: Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product. GE Healthcare will replace affected latches.
Quantity
3,053 units
Distribution
Worldwide distribution.
Lot, serial or model codes
All Lot/Serial Numbers
510(k) numbers
["K020543","K101788"]
PMA numbers
not on file
Product code
FMZ
Firm FEI
3014220748
Firm city
Wauwatosa
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register