Skip to the content

FDA Medical Device Recalls (openFDA)

Cardinal Health Inc.: Kangaroo Joey Safety Screw Spike with Flush Bag, Anti-free Flow,1000 mL, Product Code 765100

Recall number
Z-2100-2020
Recalling firm
Cardinal Health Inc.
Product
Kangaroo Joey Safety Screw Spike with Flush Bag, Anti-free Flow,1000 mL, Product Code 765100
Status
Open, Classified
Initiated
2019-12-26
Posted
not on file
Terminated
not on file
Reason
The feeding spike sets may leak at the interface of the tube and spike connector.
Root cause
Process control
Action
A Medical Device Product Advisory letter dated 12/23/19 was sent to customers. 1) INSPECT your inventory for affected product. 2) COMMUNICATE with all personnel that utilize the equipment the risk of leaks at the interface of the tube and spike connector. 3) INSPECT the feeding set for leaks after the pump has been primed and the infusion has started. " If a leak is detected the feeding set must be discarded and a new feeding set must be obtained. Inspect the new feeding set for leaks in the same manner. 4) NOTIFY any customers to whom you may have distributed or forwarded affected product about this product advisory. 5) RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652-9648 or email to gmb-fieldcorrectiveaction@cardinalhealth.com, whether or not you have affected product. Please contact the Customer Service group for any questions related to this action: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other customers888-444-5440 For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Quantity
114,731, 290 total units
Distribution
Worldwide distribution.
Lot, serial or model codes
All
510(k) numbers
["K040196","K141479","K153074"]
PMA numbers
not on file
Product code
LZH
Firm FEI
3004008697
Firm city
Dublin
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register