AGFA Corp.: 19" Barco MFCD 1219 (Touch Screen) + Low Profile Stand and Rack with LCD Arm Agfa's Computed Radiography Systems with NX2.X software are indicated for use in providing diagnostic quality images to aid the physician with diagnosis.

Recall number
Z-2100-2010
Recalling firm
AGFA Corp.
Product
19" Barco MFCD 1219 (Touch Screen) + Low Profile Stand and Rack with LCD Arm Agfa's Computed Radiography Systems with NX2.X software are indicated for use in providing diagnostic quality images to aid the physician with diagnosis.
Status
Terminated
Initiated
2010-06-30
Posted
2010-07-26
Terminated
2011-04-28
Reason
Monitor fell from the mounting bracket used to support the monitor.
Root cause
Other
Action
The firm, AGFA Healthcare sent an 'URGENT SAFETY NOTICE" dated June 30, 2010, via FED-EX to customers. The letter describes product, problem and actions to be taken by the firm and customers. AGFA will provide a safety check to the customers facility, have a service representative visit , in the near future, their site to inspect the monitor/s and attach a warning label on each monitor checked for preventative action. The customers are ask to distribute the information within their facility to all individuals who need to be aware, to complete the attached URGENT SAFETY NOTICE FEEDBACK FORM as soon as possible and return via fax at 864-421-1664. If you have any questions about this matter, please feel free to contact me via email at Debbie.norris@agfa.com or call me at 864-421-1754.
Quantity
805 units
Distribution
Worldwide distribution: USA and countries of Puerto Rico and Canada.
Lot, serial or model codes
LU8QJ000
510(k) numbers
["K071162"]
PMA numbers
not on file
Product code
MQB
Firm FEI
3001236302
Firm city
Greenville
Firm state
SC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register