Dako North America Inc.: Dako HER2 CISH pharmDx Kit, product code: SK 109 Product Usage: For in vitro diagnostic use. HER2 CISH pharmDx Kit is intended for dualcolor chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same tissue section for evaluation under bright field microscopy. The CISH procedure is automated using Dako Autostainer instruments. HER2 CISH pharmDx Kit is indicated as an aid in the assessment of patients for whom Herceptin (trastuzumab) treatment is being considered. Results from the HER2 CISH pharmDx Kit are intended for use as an adjunct to the clinicopathologic information currently used for estimating prognosis in stage II, node-positive breast cancer patients.

Recall number
Z-2099-2016
Recalling firm
Dako North America Inc.
Product
Dako HER2 CISH pharmDx Kit, product code: SK 109 Product Usage: For in vitro diagnostic use. HER2 CISH pharmDx Kit is intended for dualcolor chromogenic visualization of signals achieved with directly labeled in situ hybridization probes targeting the HER2 gene and centromeric region of chromosome 17. The kit is designed to quantitatively determine HER2 gene status in formalin-fixed, paraffin-embedded breast cancer tissue specimens. Red and blue chromogenic signals are generated on the same tissue section for evaluation under bright field microscopy. The CISH procedure is automated using Dako Autostainer instruments. HER2 CISH pharmDx Kit is indicated as an aid in the assessment of patients for whom Herceptin (trastuzumab) treatment is being considered. Results from the HER2 CISH pharmDx Kit are intended for use as an adjunct to the clinicopathologic information currently used for estimating prognosis in stage II, node-positive breast cancer patients.
Status
Terminated
Initiated
2016-06-10
Posted
not on file
Terminated
2016-09-02
Reason
An incorrect HER2 probe concentration was used when producing the probe component (Vial 3) in this kit lot. This has resulted in the HER2 probe concentration being too low.
Root cause
Process control
Action
Dako sent a recall notification dated June 10, 2016 to all customers. The letter identified the affected product, problem, the investigation findings, the affected results, and the actions to be taken. Customers are instructed to complete and return the enclosed device recall form to Dako QA Vigilance by Dako.dkvigilance@agilent.com. Customers with questions regarding the notification, are instructed to contact their sales representative.
Quantity
31 kits
Distribution
Worldwide Distribution - US Nationwide in the states of: PA, CA, AL, NC and countries of Austria, Bosnia and Herzegovina, Belgium, Bulgaria, Brazil, Canada, Switzerland, Germany, Spain, France, Greece, Italy, Macedonia, Qatar, Romania, Serbia, Sweden, and Turkey.
Lot, serial or model codes
20031683
510(k) numbers
not on file
PMA numbers
not on file
Product code
NYQ
Firm FEI
2022180
Firm city
Carpinteria
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register