FDA Medical Device Recalls (openFDA)
Stryker Instruments Div. of Stryker Corporation: 0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover 0702-002-000, Neptune 2 Rover Ultra (230V) 0702-002-000ES, Neptune 2 Rover (230V) Spanish 0702-002-000FI, Neptune 2 Ultra Rover (230V) Finnish 0702-002-000IT, Neptune 2 Rover (230V) Italian 0702-002-000R, Repair Neptune 2 Rover Ultra (230V) Product Usage: The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor s offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electro cautery or laser devices. Contraindication: connection of the Neptune 2 Waste Management System directly to chest tubes or other closed wound drains. The Stryker Neptune 2 Rover is a mobile unit used to suction, collect fluid waste, and evacuate surgical smoke from a surgical site during operation. Fluid waste is removed from the surgical site through suction tubing connected to inlet ports of disposable manifolds installed in the rover. The Diverter Valve Assembly consists of two diverter valves with three hoses, one inlet and two outlets. The sole purpose of the Diverter Valve Assembly is to direct/divert water from the fresh water inlet to either the large canister or small canister via the pre-fill tank.
- Recall number
- Z-2099-2015
- Recalling firm
- Stryker Instruments Div. of Stryker Corporation
- Product
- 0702-001-350Q, Diverter Valve Assembly 0702-001-000, Neptune 2 Rover Ultra (120V) 0702-001-000R, Repair Neptune 2 Rover 0702-002-000, Neptune 2 Rover Ultra (230V) 0702-002-000ES, Neptune 2 Rover (230V) Spanish 0702-002-000FI, Neptune 2 Ultra Rover (230V) Finnish 0702-002-000IT, Neptune 2 Rover (230V) Italian 0702-002-000R, Repair Neptune 2 Rover Ultra (230V) Product Usage: The Neptune 2 Waste Management System is intended to be used in the operating room, pathology, surgical centers, and doctor s offices to collect and dispose of surgical fluid waste as well as collect smoke generated from electro cautery or laser devices. Contraindication: connection of the Neptune 2 Waste Management System directly to chest tubes or other closed wound drains. The Stryker Neptune 2 Rover is a mobile unit used to suction, collect fluid waste, and evacuate surgical smoke from a surgical site during operation. Fluid waste is removed from the surgical site through suction tubing connected to inlet ports of disposable manifolds installed in the rover. The Diverter Valve Assembly consists of two diverter valves with three hoses, one inlet and two outlets. The sole purpose of the Diverter Valve Assembly is to direct/divert water from the fresh water inlet to either the large canister or small canister via the pre-fill tank.
- Status
- Terminated
- Initiated
- 2015-05-22
- Posted
- 2015-07-17
- Terminated
- 2016-09-12
- Reason
- Stryker Instruments is initiating a recall of the Neptune 2 Waste Management System due to complaints related to installation failures of the Neptune 2 diverter valve assembly during docking.
- Root cause
- Nonconforming Material/Component
- Action
- Stryker sent an Urgent Medical Device Notification distributed customer notices on 06/09/2015 via certified mail. Customers were asked to review the notice, complete and return the response card, and forward the notice in the event the recalled device was further distributed. Customers were asked to fax the completed Business Reply Form to Stryker Instruments Regulatory Department, 866-521-2762, or scan and email a copy to kellyjo.davis@stryker.com, and to please continue to rotate all Neptune2 Rovers. If Rovers are removed from circulation for more than 40 days, it is recommended that you contact Stryker Repair if any issue is experienced putting the Rover back into use.
- Quantity
- 9,245
- Distribution
- Worldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN. KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NY, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of: Canada, Europe, Asia Pacific, EEMEA.
- Lot, serial or model codes
- Serial Numbers ranging from 0803118093 through 1505400223; Distribution or Repair dates: November 1, 2011 to March 31, 2015
- 510(k) numbers
- ["K132671"]
- PMA numbers
- not on file
- Product code
- JCX
- Firm FEI
- 1811755
- Firm city
- Portage
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28