Skip to the content

FDA Medical Device Recalls (openFDA)

Pro-Dex Inc: KLS Martin Battery Pack, REF: KLS BP 001, and KLS BP 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver

Recall number
Z-2098-2023
Recalling firm
Pro-Dex Inc
Product
KLS Martin Battery Pack, REF: KLS BP 001, and KLS BP 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver
Status
Open, Classified
Initiated
2023-02-15
Posted
2023-07-05
Terminated
not on file
Reason
Battery packs used in surgical procedures where a surgeon uses a powered screwdriver and a right-angle attachment have switched manufacturing and expiration dates on sterile pouch labels, so batteries will appear to be prematurely expired.
Root cause
Error in labeling
Action
On 2/15/23, a recall notice was sent to a customer asking them to cease distribution and return affected product to the recalling firm. Questions can be directed to 949-769-3200.
Quantity
2000
Distribution
US: FL
Lot, serial or model codes
REF: KLS-BP-001/KLS-BP-040, UDI-DI: 00862377000140, Lot: K0M6U, Expiration: 8/31/2024
510(k) numbers
not on file
PMA numbers
not on file
Product code
GEY
Firm FEI
2081135
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register