FDA Medical Device Recalls (openFDA)
Bausch & Lomb Surgical, Inc.: SoftPort Easy-Load Lens Delivery System Model: EZ-28V Is used for the folding and delivering of the LI61AO, LI61AOV, or LI61SE 3-piece lens into the eye. The SofPort Easy-Load Lens Delivery System consists of a syringe shaped body and tip with a plunger, drawer, and haptic puller. It is a sterile, disposable plastic device designed for single use only.
- Recall number
- Z-2098-2017
- Recalling firm
- Bausch & Lomb Surgical, Inc.
- Product
- SoftPort Easy-Load Lens Delivery System Model: EZ-28V Is used for the folding and delivering of the LI61AO, LI61AOV, or LI61SE 3-piece lens into the eye. The SofPort Easy-Load Lens Delivery System consists of a syringe shaped body and tip with a plunger, drawer, and haptic puller. It is a sterile, disposable plastic device designed for single use only.
- Status
- Terminated
- Initiated
- 2017-04-05
- Posted
- not on file
- Terminated
- 2019-12-11
- Reason
- Complaints concerning kinked haptics that were occurring during lens delivery.
- Root cause
- Device Design
- Action
- Bausch & Lomb sent a Medical Device Recall letter dated April 18, 2017 to all their consignee informing them of the recall. Consignees were instructed to review their inventory and quarantine any unused product and return to Stericycle, the firm conducting this voluntary product recall on behalf of Bausch & Lomb. Contact Stericycle to arrange for a return at 1-877-469-8081. For replacement devices call 1-800-338-2020. Customers with questions were instructed to call 1-877-479-8081 or BauschandLomb5410@Stericycle.com. QUARANTINE PRODUCT AND RETURN TO BAUSCH + LOMB According to our records, your facility may have a supply of SofPort injection devices from the lots specified in this voluntary recall. We ask that you please quarantine any unused boxes (full and partial) and take the following steps to return the SofPort injection devices included in this voluntary recall to Stericycle, the firm conducting this voluntary product recall on behalf of Bausch + Lomb: 1. Please review your inventory and hold all unused (full and partial) boxes of SofPort injection devices. An example of the product label is provided below. 2. Complete the enclosed Medical Device Acknowledgement Form and return it to Stericycle. 3. Contact Stericycle to arrange for a return of the identified product in your facility at 1-877-479-8081. 4. Contact the Surgical Customer Service team to order replacement SofPort injection devices at 1-800- 338-2020. If you have questions or would like additional information, please call Stericycle, the firm conducting this voluntary recall on behalf of Bausch + Lomb, at 1-877-479-8081 or BauschandLomb5410@Stericycle.com.
- Quantity
- not on file
- Distribution
- Nationwide Distribution to AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico
- Lot, serial or model codes
- Lot Number: H671901 Expiration Date: Dec-17, Lot Number: H672001 Expiration Date: Nov-17, Lot Number: H672602 Expiration Date: Dec-17, Lot Number: H674001 Expiration Date: Dec-17, Lot Number: H674401 Expiration Date: Nov-17, Lot Number: H674801 Expiration Date: Dec-17, Lot Number: H675401 Expiration Date: Dec-17, Lot Number: H740301 Expiration Date: Dec-17, and Lot Number: H743201 Expiration Date: Feb-18.
- 510(k) numbers
- ["K970727"]
- PMA numbers
- not on file
- Product code
- MSS
- Firm FEI
- 1119279
- Firm city
- Clearwater
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28