Abbott Medical Optics Inc. (AMO): AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (Model Number NGP0020) Used with the AMO Whitestar Signature System. For ophthalmic use.
- Recall number
- Z-2096-2017
- Recalling firm
- Abbott Medical Optics Inc. (AMO)
- Product
- AMO 25GA Vitrectomy Cutter (Model Number NGP0025) and AMO 20GA Vitrectomy Cutter and Irrigation Sleeve (Model Number NGP0020) Used with the AMO Whitestar Signature System. For ophthalmic use.
- Status
- Terminated
- Initiated
- 2017-03-30
- Posted
- 2017-05-16
- Terminated
- 2017-08-23
- Reason
- Packaging error. 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package.
- Root cause
- Packaging process control
- Action
- An Urgent Field Safety notice was sent on 3/30/17 to inform customers that AMO has voluntarily initiated this Action because a possibility exists that, due to a packaging error, 20GA Vitrectomy Cutters may be found in a 25GA package and 25GA Vitrectomy Cutters may be found in a 20GA package. Use of a Vitrectomy Cutter that is a different size than expected could lead to the need to alter the surgical technique, including conjunctival dissection, incision enlargement and scleral sutures.
- Quantity
- 360 units total
- Distribution
- Worldwide distribution to SA, TR, AZ, NL, PT, SG, AU, JP, and HK.
- Lot, serial or model codes
- Part Number NGP0025 Lot CA06182 (72 units) Part Number NGP0020 Lot CA06180 (288 units)
- 510(k) numbers
- ["K081681"]
- PMA numbers
- not on file
- Product code
- MLZ
- Firm FEI
- 3003843509
- Firm city
- Santa Ana
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28