Skip to the content

FDA Medical Device Recalls (openFDA)

Custom Healthcare Systems, Inc.: Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case. Packaged for Physician use only.

Recall number
Z-2096-2015
Recalling firm
Custom Healthcare Systems, Inc.
Product
Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case. Packaged for Physician use only.
Status
Terminated
Initiated
2014-11-05
Posted
2015-07-17
Terminated
2016-01-12
Reason
An INCORRECT item number is packed within the OUTSIDE "PIGGYBACKED" PACKAGE. The interior sealed STERILE KIT is correct and NO adverse issues affect that part of the item. The NON-STERILE OUTSIDE "PIGGYBACKED" PACKAGE mistakenly contains ITEM #8881579121 - Syringe prefilled, Saline Flush, 09% Sodium Chloride and is a NONSTERILE PRODUCT that CANNOT BE DROPPED INTO THE STERILE FIELD.
Root cause
Mixed-up of materials/components
Action
Custom Healthcare System sent an Urgent Product Recall letter dated November 5, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. PLEASE CHECK YOUR INVENTORY AND QUARANTINE ANY AFFECTED PRODUCT MATCHING THE DESCRIPTIONS ABOVE. Call 804 (421-5959) with regards to returning the affected products. Return Authorizations will be issued and product pickup and return will be scheduled.
Quantity
3 Cs-Lot#92414 , 25 Cs-Lot#100214, 24 Cs-Lot#100914
Distribution
US Distribution to CT.
Lot, serial or model codes
Lot#92414 Exp Date 11/1/15, Lot#100214 Exp Date 11/1/15 Lot#100914 Exp Date 11/1/15
510(k) numbers
not on file
PMA numbers
not on file
Product code
GCB
Firm FEI
1121091
Firm city
Richmond
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register