FDA Medical Device Recalls (openFDA)
CareFusion 303, Inc.: BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191, 111-192, 111-193, 111-194, 111-195, 111-196, 111-197, 111-198; BD Pyxis CII Safe, Software: V8, 111-199-01, 111-200-01, 111-201-01, 111-202-01, 111-203-01, 111-204-01, 111-205-01, 111-206-01, 107-245-01, 107-246-01; BD Pyxis CII Safe, Software: V9, REF: 107-252-01, 111-215, 111-216, 111-219, 111-221. BD Pyxis CII Safe, Software: ES, REF: 107-254-01, 107-255-01, 1116-00, 1117-00; BD Pyxis CII Safe, Software: N/A, REF: 111-113, 111-114, 111-115, 111-160-201, 111-217, 111-218, 111-220, 111-222; Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025. Used with BD Pyxis CII Safe System: v7.x for WES7 User Guide, DIR 10000145747/ v8.1 User Guide, DIR 10000277635/v9.0 User Guide, DIR 10000354270/ ES User Guide v1.8, DIR 10000418517
- Recall number
- Z-2094-2023
- Recalling firm
- CareFusion 303, Inc.
- Product
- BD Pyxis CII Safe, Software: V7, REF: 107-96, 107-97, 107-97-201; BD Pyxis CII Safe, Software: V7, REF: 107-223, 107-224, 111-191, 111-192, 111-193, 111-194, 111-195, 111-196, 111-197, 111-198; BD Pyxis CII Safe, Software: V8, 111-199-01, 111-200-01, 111-201-01, 111-202-01, 111-203-01, 111-204-01, 111-205-01, 111-206-01, 107-245-01, 107-246-01; BD Pyxis CII Safe, Software: V9, REF: 107-252-01, 111-215, 111-216, 111-219, 111-221. BD Pyxis CII Safe, Software: ES, REF: 107-254-01, 107-255-01, 1116-00, 1117-00; BD Pyxis CII Safe, Software: N/A, REF: 111-113, 111-114, 111-115, 111-160-201, 111-217, 111-218, 111-220, 111-222; Tripp Lite SUPER6TEL Surge Suppressor, REF: CN 200-025. Used with BD Pyxis CII Safe System: v7.x for WES7 User Guide, DIR 10000145747/ v8.1 User Guide, DIR 10000277635/v9.0 User Guide, DIR 10000354270/ ES User Guide v1.8, DIR 10000418517
- Status
- Open, Classified
- Initiated
- 2023-05-22
- Posted
- 2023-07-05
- Terminated
- not on file
- Reason
- Automated dispensing cabinet devices plugged into any multi-socket outlet power strips may present unacceptable levels of current leakage, which could expose users to system downtime and/or potential electric shock.
- Root cause
- Labeling design
- Action
- On 5/22/23, correction notices were sent to distributors, directors of pharmacy/biomedical engineering/risk management who were asked to take the following actions: 1) Always select an unused, hospital grade AC outlet for the connection. Do not use an extension cord or a multiple-socket outlet strip between the BD Pyxis CII Safe equipment power cord and wall outlet. 2) If the affected device is currently plugged into a multiple socket outlet, follow the shutdown instructions below: a) On the Main Menu screen, tap Exit. b) On the Windows taskbar, tap Start > Shut Down. c) Select Shut down and tap OK. d) After the station has shut down completely, move the power switch to the off position. e) Disconnect the power cord 3) Disseminate the correction notice within your facility network to ensure they are aware of the risks. 4) Complete and return the response form to BDRC43@bd.com 5) Customers with questions can call 1.800.727.6102 or 1.858.617.2000. Emails can also be sent to: Customer Self Service Portal: https://service.carefusion.com and productcomplaints@bd.com In addition, distributors were asked to identify all customers within their distribution network that purchased any affected product as defined in the notification. Provide a copy of the included customer letter to all customers to advise them of the field action notification on BD s behalf. On 10/31/23, amended correction notices were distributed informing additional customers of additional affected devices. In addition, customers were asked not use any MSO with BD Pyxis CIISafe devices or CIISafe accessories. If affected devices are plugged into multiple socket outlets, the firm is asking the customer to now create a case using the customer Self Service Portal, or contact BD Customer Support, 1.866.583.8783 8AM-5PM CST, to schedule a controlled shutdown of the device.
- Quantity
- 9,906
- Distribution
- US Nationwide-Worldwide Distribution: CO, MS, HI, FL, GA, CA, ND, GU, DE, MO, DC, SC, VA, AK, TX, NY, MN, PA, MA, OH, WI, NC, KY, KS, IL, MD, NE, MI, IA, WA, NM, LA, OK, AZ, AR, IN, WY, AL, NJ, SD, NV, TN, OR, WV, ME, ID, CT, UT, RI, MT, PR, NH, VT, AS OUS: GU, KW, AE, SA, NZ, CA, BE ,GB, IT, JP, KR, DE, AU, SG, BH, JO, QA, ES, AT, NL, OM, BM, AT
- Lot, serial or model codes
- UDI-DI: 10885403512520; Devices Manufactured May 04, 2006 - September 26, 2023
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BRY
- Firm FEI
- 2016493
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28