Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Sarns D4 Aortic Arch Cannulae 6.0mm, 7mm & 8mm angled and straight tip with or without side holes and/or luer Product Usage: The Sarns Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.

Recall number
Z-2094-2014
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns D4 Aortic Arch Cannulae 6.0mm, 7mm & 8mm angled and straight tip with or without side holes and/or luer Product Usage: The Sarns Aortic Cannula is indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.
Status
Terminated
Initiated
2014-05-23
Posted
2014-07-23
Terminated
2015-02-24
Reason
During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications
Root cause
Process control
Action
Terumo sent an Urgent Medical Device Recall letters dated June 27, 2014 to their customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to 1. Review this Medical Device Recall notice, 2. Assure that all users in your institution have received notice of this issue and remove any affected units from use, 3. Fill out the Product Tracking Sheet (enclosed), indicating the number of affected units at your institution and fax to 1-877-907-7509 or email to terumo4967@stericycle.com. 4. Complete the Packing Slip (enclosed) and follow the packing instructions to return the affected product, and 5. Terumo CVS will issue credit once the affected units are returned. For questions call Terumo CVS c/o Stericycle 1-866-629-6181.
Quantity
7,460 units
Distribution
Worldwide Distribution - US (nationwide) in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV & WY and the countries of : Australia, Belgium, Canada, Chile, Japan, Malaysia, Mexico, Singapore, Taiwan, Thailand, United Arab Emirates (UAE) & United Kingdom.
Lot, serial or model codes
Part No. 165269, Lot #: 0649648, 0652316, 0653402, 0659824, 0678725, 0682564; Part No. 9482, Lot #: 0653227, 0656440, 0673777; Part No. 4344, Lot #: 0645166, 0653856, 0673024; Part No. 9485, Lot #: 654068; Part No. 9486, Lot #: 649867; Part No. 4400, Lot #: 0650491, 0659436, 0669133, 0674461; Part No. 165267, Lot #: 0654562, 0672209, 0676284, 0678213; Part No. 165266, Lot #: 0658973, 0661133, 0667798, 0675018, 0678084; Part No. 9483, Lot #: 0649709, 0653442, 0655043; Part No. 4460, Lot #: 0667803, 0675761, 0677655; Part No. 165765, Lot #: 677007; Part No. 165265, Lot #: 0652644, 0659820, 0674746, 0683956; Part No. 165264, Lot #: 0646481, 0650489, 0653235, 0668894, 0676282, 0681876, 0683955 & Part No. 9484, Lot #: 0649158, 0655040, 0660676.
510(k) numbers
["K874896"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register